Inspiratory response and side-effects to rapid bilateral magnetic phrenic nerve stimulation using differently shaped coils: implications for stimulation-assisted mechanical ventilation

Abstract

Background

Rapid magnetic stimulation (RMS) of the phrenic nerves may serve to attenuate diaphragm atrophy during mechanical ventilation. With different coil shapes and stimulation location, inspiratory responses and side-effects may differ. This study aimed to compare the inspiratory and sensory responses of three different RMS-coils either used bilaterally on the neck or on the chest, and to determine if ventilation over 10 min can be achieved without muscle fatigue and coils overheating.

Methods

Healthy participants underwent bilateral anterior 1-s RMS on the neck (RMSBAMPS) (N = 14) with three different pairs of magnetic coils (parabolic, D-shape, butterfly) at 15, 20, 25 and 30 Hz stimulator-frequency and 20% stimulator-output with + 10% increments. The D-shape coil with individual optimal stimulation settings was then used to ventilate participants (N = 11) for up to 10 min. Anterior RMS on the chest (RMSaMS) (N = 8) was conducted on an optional visit. Airflow was assessed via pneumotach and transdiaphragmatic pressure via oesophageal and gastric balloon catheters. Perception of air hunger, pain, discomfort and paresthesia were measured with a numerical scale.

Results

Inspiration was induced via RMSBAMPS in 86% of participants with all coils and via RMSaMS in only one participant with the parabolic coil. All coils produced similar inspiratory and sensory responses during RMSBAMPS with the butterfly coil needing higher stimulator-output, which resulted in significantly larger discomfort ratings at maximal inspiratory responses. Ten of 11 participants achieved 10 min of ventilation without decreases in minute ventilation (15.7 ± 4.6 L/min).

Conclusions

RMSBAMPS was more effective than RMSaMS, and could temporarily ventilate humans seemingly without development of muscular fatigue.

Trial registration This study was registered on clinicaltrials.gov (NCT04176744).

Background

The use of mechanical ventilation (MV) to replace spontaneous breathing is the gold standard during respiratory failure. However, the unphysiological positive pressure and prolonged diaphragmatic inactivity during MV can induce lung injury and diaphragm atrophy within 18 h [1]. Atrophy is associated with intramuscular changes at fiber and cellular levels [2,3,4], leading to a reduced ability to generate force and thus prolonged weaning time [5], increasing the risk of further damage and pulmonary complications.

One potential method to reduce ventilator-associated diaphragm atrophy is to activate the diaphragm via nerve-stimulation. Preliminary results in animals [6, 7] have shown that phrenic nerve stimulation (PNS) reduces MV-induced diaphragm atrophy. Furthermore, diaphragm pacing via PNS in spinal cord injured subjects can replace MV to induce resting breathing and preserve diaphragm function [8]. Intravenously inserted stimulation catheters have also been tested [9] and were recently shown to attenuate the decrease in inspiratory muscle strength, but to not decrease weaning time in difficult-to-wean patients [10]. However, any catheter poses a risk for infection and thus, non-invasive external PNS may serve as a favorable alternative.

Magnetic stimulation is one method for non-invasive PNS via a variety of coils and stimulation locations such as over the cervical spine (CMS) [11], bilaterally and anterolaterally on the neck (BAMPS) [12], or anteriorly on the chest (aMS) [13]. Two research groups have explored the use of rapid magnetic stimulation (RMS) in the context of diaphragm stimulation [14, 15], and this technique was recently shown to not negatively interact with intensive care unit (ICU) equipment [16]. RMS may therefore serve as an ideal candidate for non-invasive treatment. Sander et al. [14] showed that rapid bilateral anterior magnetic phrenic nerve stimulation (RMSBAMPS) can produce strong enough diaphragmatic contractions to induce short-term ventilation in healthy subjects for no longer than 5 min due to “technical prerequisites.” Adler et al. [15] sought to determine the optimal combination of stimulation-frequency and stimulator-output for tolerable rapid CMS (RMSCMS) in healthy humans, but they were unable to induce inspiratory flow despite sufficient diaphragm contraction.

With new developments of coils and cooling techniques, the present study aimed to compare inspiratory and sensory responses to RMSBAMPS when using different combinations of stimulation-frequency and stimulator-output with three differently-shaped coils, and to investigate whether a continuous series of RMSBAMPS could sustain ventilation at a constant level for 10 min. Additionally, rapid anterior stimulation on the chest (RMSaMS) was tested, as this location may offer better nerve access.

Methods

Ethical approval

The study was approved by the Cantonal Ethics Committee of Zurich (Project ID 2019-01990) and registered on clinicaltrials.gov (NCT04176744). The study conformed with the Declaration of Helsinki.

Experimental design

The study took place over 2–3 study visits. Participants abstained from intense exercise 48 h before, and from any type of exercise 24 h before each study visit, and from consumption of alcohol and caffeinated food or drinks on study days prior to testing. During visit 1, participants underwent lung function and respiratory muscle testing followed by a single-train RMSBAMPS protocol consisting of 1-s RMS trains of the phrenic nerves. The single-train RMSBAMPS protocol was conducted with three commercially available coils using various combinations of stimulator-output (% of maximum) and stimulation-frequency. During visit 2, participants had their body composition assessed prior to undergoing a series of consecutive trains of RMSBAMPS to induce ventilation in 1-min blocks, using a cooled D-shape coil, for 10 min. During an optional visit 3, participants underwent the same protocol as during visit 1, except that RMSBAMPS was replaced by RMSaMS performed on the chest. Cardiorespiratory variables, surrogate measures of respiratory muscle activation, subjective perceptions and measures of side-effects were monitored throughout all visits.

Participants

Of the 15 participants originally recruited for the study, one participant withdrew due to an initial misunderstanding of study requirements. Thus, data of 14 participants (9M:5F) are presented. Due to the COVID-19 pandemic, only five participants (4M:1F) who had completed visit 2 returned for the optional study visit, therefore three additional participants (3F) were recruited at the end of the study to only undergo the optional visit 3 (RMSaMS). Participant characteristics including anthropometrics, body composition, lung function and respiratory muscle strength are given in Table 1.

Table 1 Participant characteristics

Lung function, respiratory muscle strength, body composition

In all participants standard lung function, respiratory muscle strength, and body composition was assessed. Lung function tests were performed using a commercially available testing system and body box (Quark PFT & Q-Box, Cosmed, Rome, Italy) according to current guidelines [17]. Respiratory muscle strength was assessed via maximal inspiratory and expiratory maneuvers (from residual volume and total lung capacity, respectively) alternating every three maneuvers using a respiratory pressure meter (RP Check, MD Diagnostics LTD., Kent, United Kingdom). Percentage of predicted values were calculated for each participant using specific reference equations for lung function [18] and respiratory muscle strength [19]. Body composition was measured via a dual-energy X-ray absorptiometry (DXA) scan (lunar iDXA densitometer, GE Healthcare, Madison, WI, USA) in order to determine potential relationships between select anthropometric variables and VT in response to single-train RMSBAMPS.

Cardiorespiratory response to RMS

Participants were instrumented with a mouthpiece connected to a calibrated pneumotachometer (Series 3813, Hans Rudolph, Shawnee, KS, USA) for continuous measurement of flow. End-tidal partial pressure of carbon dioxide (PETCO2) and peripheral blood oxygen saturation (SPO2) were measured using a patient monitor (Cardiocap/5, Datex-Ohmeda, Madison, WI, USA). On all visits, participants had the option of inserting two balloon catheters (Adult Esophageal Balloon Catheters 47-9005, Cooper Surgical, Trumbull, CT, USA) that were connected to calibrated differential pressure transducers (DP45, Validyne Engineering, Northbridge, CA, USA) to record transdiaphragmatic pressure (Pdi). Briefly, both balloon catheters were inserted through the nares, following numbing with local anesthetic spray (Xylocain Spray 10%, Aspen Pharma Schweiz GmbH, Baar, Switzerland), and both balloons were first placed in the stomach. One balloon remained in the stomach to measure gastric pressure (Pga). The second balloon was withdrawn in 1 cm increments until a negative deflection was detected during a sniff maneuver, followed by the balloon being withdrawn an additional 10 cm to ensure it was completely removed from the stomach in order to measure oesophageal pressure (Poes). Participants were instructed to execute a Valsalva maneuver to empty both the gastric and oesophageal balloons which were then filled with 2 ml and 1 ml of air, respectively. Final placement was adjusted, if needed, so that end-expiratory pressure during tidal breathing resulted in a Poes of 0cmH2O. Both balloons were then secured to the nose with tape to ensure the same position throughout the experiments. Pdi was calculated as the difference between Pga and Poes. All participants elected to attempt catheter insertion but only 8 of the 14 tolerated the procedure. To determine if less invasive measures of diaphragm contraction would correlate with tidal volume (VT) and Pdi, participants were also equipped with two respiratory belts (TN1132/ST, ADInstruments, Dunedin, New Zealand). One belt was placed over the naval, while one was placed along the nipple line in men and the highest possible position below the breasts in women to measure changes in abdominal and chest circumference, respectively (ΔAbdominal and ΔChest). Finally, cardiovascular changes in response to RMS were monitored with a simple 3-lead electrocardiogram connected to a bioamplifier (PowerLab 15T, Dunedin, New Zealand) and a plethysmographic finger cuff (Nexfin, Edwards Lifesciences, Amsterdam, Netherlands) for continuous measurement of heart rate (HR) and blood pressure (BP), respectively.

Side-effects in response to RMS

Participants rated their perception of pain, discomfort and paresthesia in response to each RMS setting by pointing to a visual scale ranging from 0-to-10points. 0 was anchored as “none” while 10 was anchored as “maximal” (the maximum that one could imagine). Participants were also asked specifically whether they felt dental pain. During the continuous RMSBAMPS-ventilation trial on visit 2, the perception of air hunger was also evaluated via the same 0-to-10point visual scale. Galvanic skin response (GSR), a surrogate measure for changes in one’s emotional response, was measured on the fingers using a commercially available GSR system (MLT118F GSR Electrodes & FE116 GSP Amp, ADInstruments, Dunedin, New Zealand). Lastly, the presence of upper air way collapse during RMS was quantified by counting the instances of an increase in Pdi or ΔAbdominal without concomitant flow as seen in Fig. 1B.

Fig. 1
figure 1

Example of the data analysis window during an inspiratory response and of an upper airway collapse induced by rapid bilateral anterior magnetic phrenic nerve stimulation (RMSBAMPS). A Thoracic excursion, abdominal excursion, transdiaphragmatic pressure, flow and tidal volume in response to RMSBAMPS (30 Hz stimulation-frequency, 20% stimulator-output, D-shape coil) with the glottis open. Dashed lines between the start of the stimulator signal and the end of inspiration represent the analysis window in which the mean and/or peak values were used in the analysis of select variables. B Example of an upper airway collapse during RMSBAMPS (30 Hz stimulation-frequency, 30% stimulator-output, butterfly coil) represented by an increase in transdiaphragmatic pressure without the presence of flow. Data within figure has been smoothed

Single-train RMSBAMPS (visit 1)

Single-train RMSBAMPS was conducted using two identical commercially available magnetic stimulators (MagPro X100, MagVenture, Farum, Denmark) and 3 pairs of differently-shaped magnetic coils: a parabolic-shaped coil (MMC-90, MagVenture, Farum, Denmark [max dB/dt = 30 kT/s, 280 μs]), a butterfly-shaped coil (Cool-B65, MagVenture, Farum, Denmark [max dB/dt = 36 kT/s, 290 μs]), and a D-shaped coil (Cool-D50, MagVenture, Farum, Denmark [max dB/dt = 27 kT/s, 248 μs]) (Fig. 2). The butterfly and D-shaped coils were also equipped with an active cooling unit (Coil cooler unit + high performance option, MagVenture, Farum, Denmark). Both stimulators were synchronized to stimulate at the same time upon receiving an input from an external trigger. All RMSBAMPS took place with the participant laying in a hospital bed that was tilted upright so that the torso was raised 30°. Participants’ heads were positioned within a vacuum cushion (Vacuform® 2.0 vacuum pillow 30 × 40 cm, Synmedic AG, Zurich, Switzerland) so that their necks were slightly extended. Positions of anatomical landmarks were recorded to ensure that all stimulations occurred with the participant in the same head and body position, as well as to ensure that the body position could be accurately repositioned on subsequent visits (for details also see below).

Fig. 2
figure 2

Experimental Setup. A Overview of all sensors and an example of the D-shape coil pair placed bilaterally on the neck. B An example of the butterfly coil pair placed bilaterally on the neck. C Close up of the coil and participant reflective markers, and an example of the parabolic coil pair placed bilaterally on the chest. All participants gave written consent for image use

The single-train RMSBAMPS protocol began with 3 min of resting breathing to quantify baseline measures and subsequently the placement of the first coil pair anterolaterally on the neck in the position that yielded the highest response to stimulation. The coils were initially placed in the position that yielded the highest Pdi in response to a single bilateral stimulation (Pdi,tw) or—in participants who did not tolerate catheter insertions—the highest VT in response to a 1-s train of RMSBAMPS at 20 Hz and 20–40% of stimulator-output, depending on participant tolerance. Final adjustments of coil positions were made to minimize any excessive movement of the arms and head during RMSBAMPS. The final position was secured with custom-made lever arms and recorded using the Brainsight® neuro-navigation system (BrainSight® TMS, Rogue Research Inc., Montreal, Canada) with reflective markers placed on the coils and on the participant via custom made glasses (Fig. 2). A maximum of 1 h was used to place the coils.

Following placement of the coils, trains of RMSBAMPS were systematically conducted at 15, 20, 25 and 30 Hz in a randomized order. For each frequency, the initial stimulator-output was set at 20% of maximum and increased in 10% increments until the participant no longer tolerated the increase. These 1-s trains of RMSBAMPS occurred at the end of a passive expiration with the glottis open and upon signal of the participant (Additional file 1). All stimulation-frequency and stimulator-output combinations were applied at least twice unless participants indicated they could not tolerate a second train at that setting. In the event of data contamination (ie. swallow, upper airway collapse, improper timing, etc.), additional trains of RMSBAMPS were conducted at that setting on the tester’s discretion. Following RMSBAMPS with the four tested frequencies and all tolerable stimulator-outputs, the protocol was repeated starting with coil placement of the second pair of coils and after completion of that protocol, with the third pair. Coils were tested in a randomized order. Lastly, if a participant did not show any flow response, visit 1 was repeated on another day, at least 24 h apart of the first visit 1. If no flow could be initiated on the second day, these participants were considered as non-responders.

Continuous RMSBAMPS-ventilation (visit 2)

Participants in whom flow was induced via RMSBAMPS underwent thorough familiarization with continuous RMSBAMPS-ventilation. During this familiarization, the optimal stimulation-frequency and stimulator-output combination of visit 1—using the pair of cooled coils that had yielded the largest VT at a submaximal stimulator-output (always the D-shape coil)—was tested at a respiratory frequency selected initially to match resting ventilation. If needed, adjustments to stimulator-output (± 5% increments), stimulation duration (+ 1 ms increments) and respiratory frequency (+ 1 breath-per-min increments) were made in attempt to optimally balance subjectively and objectively sufficient ventilation as well as suppression of participants' natural drive to breathe, while keeping their perception of air hunger, pain, discomfort and paresthesia tolerable throughout. As such, participants' verbal feedback was sought and taken into account to optimize their ability to complete the ten 1-min blocks of continuous RMSBAMPS-ventilation, as well as to reduce their urge for spontaneous breathing.

After familiarization, a 3-min resting breathing period was recorded followed by 1 min of ventilation. Between each train of RMSBAMPS, participants passively expired and were instructed to not initiate the next inspiration (Additional file 2). The flow signal was continuously monitored to ensure participants did not initiate breaths themselves. After each 1-min stimulation block, participants were asked to rate their perception of air hunger, pain, discomfort, and paresthesia, and whether they could tolerate a further minute of continuous RMSBAMPS. This break lasted the minimum amount of time to collect the sensory ratings. This procedure was repeated until ten 1-min blocks were completed or until participant cessation.

Single-train RMSaMS (visit 3)

The optional study visit 3 replicated the RMSBAMPS protocol from visit 1 (N = 5) with RMSaMS conducted on the chest. Coils were tested in two positions: (1) two coils placed bilaterally on either side of the chest; and (2) a single coil placed in the middle of the chest in attempt to stimulate both phrenic nerves simultaneously. A maximum of 1 h was spent to optimize the position. The three additionally recruited subjects performed a DXA scan, lung function and respiratory muscle strength tests and the same RMSaMS protocol.

Data acquisition and analysis

All physiological measurements were converted from analogue to digital with two 16-channel data acquisition systems (PowerLab 16/35, ADInstruments, Dunedin, New Zealand) and collected using LabChart Software (Version 8, ADInstruments, Dunedin, New Zealand) with a sampling frequency of 2000 Hz. VT was calculated by taking the integral of the flow measurement which was then body temperature pressure saturated corrected. Within a data analysis window that started at the beginning of RMS and ended at the absence of inspiratory flow (Fig. 1A), VT, Pdi,mean and Pdi,peak, as well as mean ΔAbdominal and ΔChest were calculated. Peak ΔGSR was analyzed outside of this window due to the latency of this signal. All responses with the same RMS setting that did not induce total upper airway collapse and where pairs of stimulations were available, were included in the analysis of inspiratory variables and averaged. Stimulations that induced total upper airway collapse were not included in the analysis of inspiratory variables, but were still quantified to determine the prevalence of this side-effect. Responses during the continuous RMSBAMPS-ventilation trial were analyzed breath by breath in the same manner as the single-train RMS and averaged over 1 min, not including the short breaks. In addition, an average of all 1-min blocks was calculated to get a total overview of the RMSBAMPS-ventilation trial. The tension-time-index of the diaphragm (TTIdi,mean) was calculated as the average pressure the diaphragm produced during inspiration compared to the maximum a participant could achieve during a maximal maneuver (Pdi,mean/Pdi,max) multiplied by the duty cycle (inspiratory duration/breath cycle). Non-responders were not included in the analysis.

A number of factors contributed to an unequal sample size within and between coils at select RMS settings (Fig. 4A) and between visits (Fig. 3). First, not all participants were able to tolerate the same maximal RMS settings within and between coils. Second, not all participants were equipped with balloon catheters during each visit. Third, one participant was not able to undergo single-train RMSBAMPS with the butterfly coil due to too much facial paresthesia. Fourth, one participant was unable to return for visit 2 due to the COVID-19 pandemic. Finally, only five of 12 participants returned for the optional visit 3, so three additional participants were recruited to undergo visit 3 only. Therefore, the number of people included in the analysis is summarized in Fig. 3.

Fig. 3
figure 3

Summary of participant recruitment and analysis. RMS: rapid magnetic stimulation. Fifteen participants were initially recruited, and one participant withdrew. As such, 14 participants were included in the analysis for single-train RMS on the neck (RMSBAMPS), which occurred on visit 1. Of those 14 participants, 8 participants were equipped with balloon catheters and two of those were non-responders. One participant with balloons did not undergo single-train RMSBAMPS with the butterfly coil due to facial paresthesia. Therefore, 12 participants (six with balloons) could be analyzed with the parabolic and D-shape coil, while 11 participants (five with balloons) could be analyzed with the butterfly coil. One participant (without balloons) was unable to return for the RMSBAMPS-ventilation trial (visit 2) due to the COVID-19 pandemic, while one participant who was equipped with balloons on visit 1 declined to undergo balloon catheter insertion on visit 2. One additional participant (without balloons) was unable to be ventilated with continuous RMSBAMPS. As such, ten participants (five with balloons) could be analyzed for RMSBAMPS-ventilation. Five participants elected to return for the optional visit in which single-train RMS was conducted on the chest, and three additional participants were recruited to undergo this visit only. Only one participant (without balloons) showed an inspiratory response to RMS on the chest with the parabolic coil only and could be analyzed

Statistics

Due to the unequal sample size with each RMS setting, a mixed effects model two-way ANOVA with repeated measures was used to determine the effect of stimulation-frequency and stimulator-output on VT, Pdi, sensory ratings and ΔGSR within each coil. Coils were compared between each other at select VT using a mixed effects model one-way ANOVA with repeated measures. Tukey’s post hoc test was used in the event of a significant effect to determine which specific coils differed. A Pearson’s correlation coefficient was used to determine within-participant relationships between ΔAbdominal and ΔChest with both VT and Pdi,mean; while a Spearman correlation coefficient determined within-participant relationships between ΔGSR and all sensory ratings during single-train RMSBAMPS. A repeated measures correlation was used to determine overall relationships between the same variables. Pearson’s correlation coefficients were also used to determine relationships between select participant anthropometric measurements (body mass index [BMI], body fat percentage, neck circumference and neck fat percentage) and VT in response to RMSBAMPS. Paired t-tests were used to compare mean continuous RMSBAMPS-ventilation trial responses with resting breathing, while a linear regression was used to determine if cardiorespiratory or sensory values significantly increased or decreased over time. All statistical analyses were conducted using Prism (v8.3.1, GraphPad Software, San Diego, CA, USA) except for the repeated measures correlation that was conducted using the RStudio’s (v1.4.1106, Boston, MA, USA) “rmcorr” package. All values are expressed as mean ± SD unless otherwise stated. P-values less than 0.05 were considered statistically significant.

Results

Effect of stimulation-frequency and stimulator-output on inspiratory responses

Inspiration in response to single-train RMSBAMPS was induced in 12 of 14 participants. Both non-responders were female (with Pdi measurement) and showed the same non-response on two visits. In general, increasing both stimulator-output and stimulation-frequency resulted in an increase in VT (N = 12) and Pdi (N = 6) as displayed in Fig. 4B and D, with larger changes with stimulator-output compared to frequency increases. Increased stimulator-output had a significant effect on VT with all coils (all P ≤ 0.0026); while an effect of frequency was only present with parabolic (P = 0.0234) and D-shape (P = 0.0405) coils. An interaction effect between stimulator-output and frequency was not present in any coil. There was a significant effect of stimulator-output on Pdi,mean (all P ≤ 0.0045) and Pdi,peak (all P ≤ 0.0022) with all coils, while frequency had an effect on both variables only when using the D-shape and parabolic coils (all P ≤ 0.0364), and no coil displayed an interaction effect between stimulator-output and frequency. Both VT and Pdi,mean showed a stronger correlation with ΔAbdominal than with ΔChest (Table 2).

Fig. 4
figure 4

Inspiratory and side-effect responses to rapid bilateral anterior magnetic phrenic nerve stimulation (RMSBAMPS) with different stimulator-output and stimulation-frequency combinations. A Number of subjects that showed an inspiratory response to RMSBAMPS that reached each tested stimulation setting; one participant did not undergo RMSBAMPS with the butterfly coil. B Tidal volume in response to RMSBAMPS (N = panel A). C Number of subjects who were equipped with balloon catheters that showed an inspiratory response to RMSBAMPS and reached each tested stimulation setting. D Mean transdiaphragmatic pressure responses to RMSBAMPS (N = panel C). EG Pain, discomfort and paresthesia in response to RMSBAMPS (N = panel A). H Change in galvanic skin response in response to RMSBAMPS (N at 20% = 10 due to technical difficulties). *, main effect of stimulator-output; , main effect of stimulation-frequency; , interaction effect (P < 0.05). Values in panels A and C are absolute count, while all other values are mean + SD

Table 2 Group and individual correlations for respiratory belts with inspiratory responses, and change in galvanic skin response with sensory responses during single-train rapid bilateral anterior magnetic phrenic nerve stimulation

Side-effects of RMSBAMPS

In general, sensory responses increased more prominently with increases in stimulator-output than increases in frequency (Fig. 4E–G). The largest stimulator-output tolerated in all stimulation-frequency and coil combinations was 20% (Fig. 4A). For pain perception, all coils showed significant effects of stimulator-output (all P ≤ 0.0002). The parabolic (P = 0.0120) and butterfly (P < 0.0001) coil showed a significant effect of frequency, and only the butterfly showed an interaction effect between stimulator-output and frequency (P = 0.0019). For discomfort, all coils had a significant effect of stimulator-output (all P < 0.0001) and frequency (all P ≤ 0.0366), with a significant interaction between the two present in the butterfly only (P = 0.0163). For paresthesia, all coils had an effect of stimulator-output (all P ≤ 0.0042), but an effect of frequency was only present with parabolic (P = 0.0112) and butterfly (P = 0.0052) coils, without interaction effects between stimulator-output and frequency with any coil. One participant was unable to undergo RMS with the butterfly coil citing too much facial paresthesia.

Changes in GSR in response to single-train RMSBAMPS are given in Fig. 4H. Two participants were excluded from analysis due to technical difficulties. An effect of stimulator-output on ΔGSR was present in all coils (all P ≤ 0.0005), an effect of frequency was only present in parabolic (P = 0.0102) and butterfly (P = 0.0499) coils, while an interaction effect between stimulator-output and frequency was only present with the butterfly (P = 0.0113). ΔGSR showed a moderate positive correlation with both pain (r = 0.55, P < 0.0001) and discomfort (r = 0.63, P < 0.0001), and a weak positive correlation with paresthesia (r = 0.38, P < 0.0001) (Table 2).

All incidents of UAC and dental pain are shown in Table 3. UAC was a common side-effect most prevalent when stimulating at the lowest frequency (15 Hz) and stimulator-output (20%). UAC was most likely to occur with the D-shape coil and least likely with the butterfly. The prevalence of dental pain (most prevalent with the butterfly coil) increased with increasing stimulator-output and occurred more often at 25 and 30 Hz compared to 15 and 20 Hz.

Table 3 Prevalence of upper airway collapse and dental pain during single-train rapid bilateral anterior magnetic phrenic nerve stimulation (RMSBAMPS)

Between coils comparisons

For any given stimulation-frequency and stimulator-output combination, the parabolic and D-shape coils achieved larger VT compared to the butterfly. With the butterfly, participants tolerated higher levels of stimulator-output achieving the same maximal VT as with the other coils at the highest tolerated stimulator-output (all P > 0.05), but with significantly larger pain (vs. parabolic + 1.2 ± 1.7 points, P = 0.0455; vs. D-shape + 1.5 ± 1.9 points, P = 0.0304) at 30 Hz. The highest achieved VT and the associated sensory ratings are presented in Fig. 5. Maximal VT did not differ between coils (parabolic 1.08 ± 0.40 L; D-shape 1.11 ± 0.34 L; butterfly 1.06 ± 0.41 L, P = 0.8037), but D-shape and butterfly coils needed more stimulator-output to achieve these volumes (parabolic = 29 ± 9%; D-shape = 37 ± 8%; butterfly = 45 ± 9%). Corresponding Pdi,mean in participants who were equipped with balloon catheters during these stimulations was 10.7 ± 5.5 (parabolic), 14.4 ± 9.7 (D-shape) and 8.8 ± 5.3 cmH2O (butterfly). During these maximal stimulations, there was a significant effect of coil on discomfort (P = 0.0471) with post-hoc tests revealing significantly higher ratings between parabolic and butterfly coils (4.9 ± 2.0 vs. 6.5 ± 2.3 points, P = 0.0462), but no difference in pain nor paresthesia. When considering only stimulations that resulted in a VT between 4-8 ml/kg body weight (Fig. 6), which reflects the target VT during MV in patients with acute respiratory distress syndrome (ARDS) [20], sensory ratings were smaller compared to stimulations where maximal VT was reached. Mean pain, discomfort and paresthesia ratings during those stimulations was 1.1 ± 1.1, 2.5 ± 1.9 and 0.8 ± 1.1 points (parabolic), 1.4 ± 1.2, 2.3 ± 1.6 and 1.5 ± 1.6 points (D-shape) and 2.0 ± 2.1, 3.3 ± 2.4 and 0.9 ± 1.4 points (butterfly). Finally, correlations between BMI, body fat percentage, neck circumference or neck fat percentage with VT in response to RMSBAMPS can be found in Additional file 3 of this manuscript. None of these variables correlated significantly in a physiologically meaningful way, when considering coil pair type, stimulation-frequency, or stimulator-output.

Fig. 5
figure 5

Highest tidal volume achieved in each coil in response to single-train rapid bilateral anterior magnetic phrenic nerve stimulation (RMSBAMPS) and associated sensory ratings with each coil. Black horizontal line in all panels represent mean values, while black dots represent individual participant values. N = 12 except for the butterfly coil data (N = 11) as one participant did not undergo RMSBAMPS with that coil. * Main effect of coil; significantly different from parabolic (P < 0.05)

Fig. 6
figure 6

Tidal volume and sensory responses to single-train rapid bilateral anterior magnetic phrenic nerve stimulation (RMSBAMPS) in which the tidal volume achieved was between 4 and 8 ml/kg of body weight. All dots represent a single data point. Black, white and grey dots represent RMSBAMPS with the parabolic, D-shape and butterfly coil, respectively. Data points are not present in participants 1 and 10 as all tidal volumes were outside of the 4–8 ml/kg body weight range (both higher and lower values achieved for both participants)

RMSBAMPS-ventilation trial

Eleven participants underwent the RMSBAMPS-ventilation trial (Fig. 7) with the D-shape coil that did not overheat. Nine completed the 10-min protocol, one completed 6 min (stopped due to discomfort; 7 points), while in one participant ventilation could not be achieved. Five participants were hyperventilated compared to resting breathing resulting in an overall slight hyperventilatory response (PETCO2 = 33.9 ± 5.0 vs. 42.0 ± 3.4 mmHg, P = 0.0011; SPO2 98 ± 1%) caused by an increase in VT (1.14 ± 0.30 vs. 0.84 ± 0.13 L, P = 0.0081), but not breathing frequency (13.9 ± 2.1 vs. 12.5 ± 3.0 breaths per minute, P = 0.1053). Both minute ventilation (\(\dot {\text{V}}_\text{E}\)) (P = 0.1562) and VT (P = 0.2157) stayed constant over time. Mean SBP (141 ± 9 vs. 122 ± 10 mmHg, P = 0.0027) and DBP (82 ± 6 vs. 72 ± 5 mmHg, P = 0.0067) were elevated compared to baseline, while only SBP significantly increased throughout the trial (P = 0.0242). HR remained at baseline levels (64 ± 8 vs. 62 ± 8 beats per minute, P = 0.1654) and did not significantly change over time (P = 0.1012). Pain (10-min average: 1.2 ± 1.1 points), discomfort (2.5 ± 1.5 points), paresthesia (1.8 ± 1.7 points) and air hunger (1.1 ± 1.7 points) did not significantly change over the 10-min duration (all P ≥ 0.3872). Participants with balloon catheters (N = 5) had a Pdi,mean of 9.4 ± 3.5 cmH2O and a TTIdi,mean of 0.02 ± 0.01 throughout the 10 min, which did not change over time (P = 0.8328 and 0.9370, respectively).

Fig. 7
figure 7

Ventilatory, sensory and cardiovascular responses to continuous rapid bilateral anterior magnetic phrenic nerve stimulation. Black data points represent the group means for each 1-min block of ventilation, while white data points represent individual participants. Each participant was stimulated with the D-shape coil. *Significant regression

RMSaMS on the chest

Eight participants underwent the single-train RMSaMS-protocol and only one participant (without balloons) showed a VT response, which could only be achieved with the parabolic coil. Inspiration was achieved with all stimulation-frequencies, but required ≥ 40% of stimulator-output. The maximal tolerated stimulator-output was 50% (15, 20, 25 Hz) and 40% (30 Hz). The range of VT was 0.11–0.30 L, while pain ranged between 5-7 points, discomfort 7-8 points and paresthesia 6-10 points.

Discussion

In the present study, RMSBAMPS could elicit flow in \(\approx\)86% of participants with parabolic, D-shape and butterfly coils, and the resultant diaphragm contractions induced VT similar or larger than during spontaneous resting breathing. The butterfly coil required the highest stimulator-output for VT similar to the other coils which resulted in larger discomfort at maximal responses. In contrast, bilateral RMSaMS could only elicit flow with the parabolic coil in one participant (13%). Finally, all but one participant (91%) reached 10 min of continuous RMSBAMPS-induced ventilation, without coils overheating and without a decrease in \(\dot {\text{V}}_\text{E}\) over time.

Single-train RMS

While all three coils produced similar inspiratory responses, the butterfly coil required the largest stimulator-output despite having the largest maximal magnetic flux. The discrepancy between responses of different coils at the same stimulator-output likely reflects the butterfly being the largest coil making it difficult to optimally place often resulting in a small space between skin and coil surface. The higher stimulator-output needed to produce similar VT also resulted in the largest discomfort at maximal responses, possibly due to the butterfly coil causing more dental pain.

Adler et al. [15] reported that 15 Hz at 65% of stimulator-output served as the optimal setting using RMSCMS to achieve the highest Pdi,peak (\(\approx\) 20 cmH2O) with sensory ratings below 3 points. Using the same criteria in the present study (pain below 3 points; ≥ 4 participants equipped with balloons), the optimal settings would be 25 Hz at 30% stimulator-output with the parabolic (Pdi,peak = 14.8 ± 4.5 cmH2O), 20 Hz at 40% with the D-shape (Pdi,peak = 19.6 ± 5.6 cmH2O), and 15 Hz at 40% with the butterfly (Pdi,peak = 8.4 ± 2.9 cmH2O) coils. These Pdi, as well as maximal Pdi, are all lower than the ones of Adler et al. [15]. However, this difference likely reflects that our participants kept their glottis open to allow airflow, while Adler et al.’s [15] kept their glottis closed during stimulation. Also, CMS is known to produce larger Pdi twitches in response to single stimuli compared to bilateral stimulations [13, 21], attributed to recruitment of accessory respiratory muscles leading to chest wall stiffening [21], that likely also applies during RMSCMS. The diaphragm activation achieved in the present study in response to RMSBAMPS reached and exceeded normal resting VT, and a level of stimulation likely able to attenuate diaphragm atrophy during MV. Recently, Sotak et al. [22] showed that percutaneous electrical PNS in ICU-patients during MV keeping the work of breathing between 0.2 and 2.0 J/L reduced the rate of the development of diaphragmatic atrophy. However, in order to maximize therapeutic effects, methods to increase the level of activation without increasing VT and thus without increasing the risk for lung injury, should be further explored.

Juxtapose to RMSBAMPS, RMSaMS only induced inspiration in one participant. As such, the technique is likely not useful for diaphragm pacing. The difference between stimulation techniques likely reflects that RMSaMS produced more movement of the shoulders, arms and torso compared to RMSBAMPS, which may have resulted in a shift in coil position (despite all efforts to avoid this) or an increased thickness of tissue between coils and phrenic nerves. The RMSaMS-responder was indeed small in stature (height = 168 cm, BMI = 19.8 kg/m2, female), but another like female (height = 165 cm, BMI = 19.6 kg/m2) did not show a flow response, possibly due to this participant’s lower tolerability of stimulator-output (40% vs. 60%). Also, flow could only be induced with the parabolic coil, possibly because this coil has a more focused point of stimulation compared to the others. In addition, VT was lower, and all sensory ratings, especially paresthesia in the arms, were elevated compared to the mean responses of all participants at matched stimulation settings during single-train RMSBAMPS (a protocol this participant did not perform).

The present study shows that a non-invasive abdominal belt may serve as a useful tool to quantify the amount of diaphragm activation during RMS given that ΔAbdominal correlated with both VT and Pdi,mean. Similarly, given that ΔGSR correlated with both pain and discomfort, it may serve as a useful measure of distress in unconscious patients (Table 2). Finally, Sander et al. [14] previously found increased V̇E in response to continuous RMSBAMPS in participants with larger BMIs and smaller neck circumferences. The \(\dot {\text{V}}_\text{E}\) achieved in the present study during RMSBAMPS-ventilation cannot be compared in a way similar to Sander et al. [14] since participants were stimulated with varying stimulation-frequencies and stimulator-outputs. However, analyses considering single-train RMSBAMPS did not detect meaningful relationships between resulting VT and either BMI, body fat percentage, neck circumference or neck fat percentage (see Additional file 3). Given the discrepancies in findings between the two studies, further investigation is needed to determine the influence body composition has on ventilatory responses to RMS.

Upper airway collapse

Similar to Sander et al. [14] (100%), RMSBAMPS successfully induced flow in \(\approx\) 86% of supine participants. In contrast, Adler et al. [15] were unsuccessful in inducing flow during RMSCMS in sitting participants when instructed to keep their glottis open, which they attributed to UAC. Stimulation technique and the differences in neck position (flexed vs. extended—known to affect upper airway dynamics and flow pattern during phrenic nerve stimulation [23]) may contribute to this discrepancy. Notably, despite flow being induced in the majority of our participants, UAC was still prominent. All but one participant experienced full or partial UAC at least once during single-train RMSBAMPS and two participants appeared to experience total UAC during all single-trains over two visits despite adjustments in stimulator-output, stimulation-frequency and coil positions. However, UAC may be an irrelevant issue in intubated and non-invasively ventilated patients given that Adler et al. [15] successfully alleviated UAC during RMSCMS when positive pressure ventilation was added. Thus, positive airway pressure may be needed to avoid closure, but this needs to be explored more systematically.

RMSBAMPS-ventilation

Continuous RMSBAMPS induced ventilation in ten of 11 participants, while one participant experienced consistent UAC, despite responding to single-train RMSBAMPS. The 10-min limit was a result of our protocol rather than equipment overheating or participant intolerance (all that achieved 10 min indicated they could go longer). Thus, with proper coil cooling, it is possible to overcome the 5-min limitation seen by Sander et al. [14].

Inducing ventilation via RMSBAMPS appears to be a tolerable technique given the majority (90%) of participants with continuous ventilation were able to complete the 10-min protocol. In fact, mean perception of pain, discomfort and paresthesia were rated below 3 points, however, one participant ended the trial early citing discomfort resulting from excessive shoulder movement on one side. This resulted from the coil moving out of the optimal position during the trial and reflects the importance of maintaining the same position between consecutive stimulations. All but two participants required slightly higher than 1-s stimulation duration (up to 1.3 s) for optimal comfort, and six participants required a higher respiratory frequency compared to their resting breathing to suppress their urge to breathe. These adjustments occurred during familiarization based on their feedback and resulted in a slight hyperventilation in five cases. Perception of air hunger was rated ≤ 2 points at all times in all but two participants. It should be noted, however, that a brief break was taken between each minute to assess participant’s sensory ratings in which they resumed spontaneous breathing which may have contributed to reduced sensory ratings. Thus, the efficacy of ventilation without breaks and a longer than 10 min still needs further exploration.

The mean \(\dot {\text{V}}_\text{E}\) achieved in the present study was similar to the 14.0 L/min by Sander et al. [14] with 25 Hz at 40% stimulator-output (two MagStim butterfly coils) in 10 healthy volunteers. Their ventilation, however, surpassed the present when stimulator-output was increased to 50% (18.6 L/min), but this increase in stimulator-output resulted in only three participants undergoing ventilation due to a largely increased perception of pain and discomfort, similar to our single-train RMSBAMPS responses with maximal stimulator-output. The present study optimized RMS settings to each participant and was 20 (N = 3), 25 (N = 5) or 30 (N = 2) Hz with a mean stimulator-output of 27% (range = 20–40%). Stimulation parameters and thus VT likely exceeded the levels that would be used or needed in mechanically ventilated patients in order to prevent lung injury. In any case, it is encouraging to note that in the present study with an excessive ventilation present, mean TTIdi was only 0.02 which is well below the 0.15 fatigue-inducing threshold [24]. As such, RMSBAMPS seems unlikely to cause fatigue-inducing contractions, at least in our group of young healthy participants, and as previously shown with CMS [15]. To which extent muscle fatigue could play a role in various patient groups needs, however, further exploration.

Finally, while Sander et al. [14] reported no cardiovascular changes (unspecified), the present study showed an increased BP during RMSBAMPS-ventilation, but without changes in HR. Although the increase in BP did not reach an unsafe level in these young subjects and stimulator-output was likely higher than what would be used in patients, it is still suggested that the cardiovascular system is closely monitored during continuous RMS.

Limitations

A few limitations remain. First, although flow traces were monitored to guarantee participants did not initiate inspiration, volitional assistance cannot be excluded with certainty. Second, the COVID-19 pandemic impacted the number of participants that returned for visit 2 and the optional visit. Third, not all participants tolerated balloon catheters resulting in a reduced sample with Pdi data, two being non-responders.

Perspective

Although the coils tested could provide sufficient inspirations, for use in an ICU setting, coils would need to be developed that are easier to handle (lighter, smaller and more maneuverable for precise repositioning), specifically designed for phrenic nerve stimulation, and ideally being run by one stimulator only. To optimize synchronisation between magnetic stimulation and mechanical ventilation, a trigger to initiate stimulation in response to either pressure or flow changes of the ventilator would be needed. Finally, the ventilator would need to adjust its flow or pressure support in real time with respect to any superimposed phrenic nerve stimulation to ensure patients are ventilated within safe volume and pressure limit.

Conclusions

RMSBAMPS, but not RMSaMS, can induce strong enough diaphragmatic contractions to ventilate healthy humans for 10 min. However, we currently do not recommend replacing MV with RMS, but rather to use RMS to assist MV in order to potentially reduce ventilator-induced diaphragm atrophy. For most effective clinical use, newly designed equipment should be less bulky and optimized for PNS with stimulators automatically adjusting timing and output according to feedback from ventilation and non-desired side-effects.

Availability of data and materials

Data can be made available upon reasonable request to the corresponding author.

Abbreviations

aMS:

Anterior magnetic stimulation

BMI:

Body mass index

CMS:

Cervical magnetic stimulation

DBP:

Diastolic blood pressure

DXA:

Dual X-ray absorptiometry

GSR:

Galvanic skin response

HR:

Heart rate

Hz:

Hertz

ICU:

Intensive care unit

MV:

Mechanical ventilation

Pdi :

Transdiaphragmatic pressure

PETCO2 :

End tidal partial pressure of carbon dioxide

PNS:

Phrenic nerve stimulation

RMS:

Rapid magnetic stimulation

RMSaMS :

Rapid anterior magnetic stimulation

RMSCMS :

Rapid cervical magnetic stimulation

RMSBAMPS :

Rapid bilateral anterior magnetic phrenic nerve stimulation

SPO2 :

Peripheral oxygen saturation

SBP:

Systolic blood pressure

TTIdi :

Diaphragm tension-time index

UAC:

Upper airway collapse

\(\dot {\text{V}}_\text{E}\) :

Minute ventilation

VT :

Tidal volume

References

  1. Levine S, Nguyen T, Taylor N, Friscia ME, Budak MT, Rothenberg P, Zhu J, Sachdeva R, Sonnad S, Kaiser LR, et al. Rapid disuse atrophy of diaphragm fibers in mechanically ventilated humans. N Engl J Med. 2008;358:1327–35.

    Article  CAS  Google Scholar 

  2. Jaber S, Petrof BJ, Jung B, Chanques G, Berthet JP, Rabuel C, Bouyabrine H, Courouble P, Koechlin-Ramonatxo C, Sebbane M, et al. Rapidly progressive diaphragmatic weakness and injury during mechanical ventilation in humans. Am J Respir Crit Care Med. 2011;183:364–71.

    Article  CAS  Google Scholar 

  3. Picard M, Jung B, Liang F, Azuelos I, Hussain S, Goldberg P, Godin R, Danialou G, Chaturvedi R, Rygiel K, et al. Mitochondrial dysfunction and lipid accumulation in the human diaphragm during mechanical ventilation. Am J Respir Crit Care Med. 2012;186:1140–9.

    Article  CAS  Google Scholar 

  4. Berger D, Bloechlinger S, von Haehling S, Doehner W, Takala J, Z’Graggen WJ, Schefold JC. Dysfunction of respiratory muscles in critically ill patients on the intensive care unit. J Cachexia Sarcopenia Muscle. 2016;7:403–12.

    Article  Google Scholar 

  5. Jung B, Moury PH, Mahul M, de Jong A, Galia F, Prades A, Albaladejo P, Chanques G, Molinari N, Jaber S. Diaphragmatic dysfunction in patients with ICU-acquired weakness and its impact on extubation failure. Intensive Care Med. 2016;42:853–61.

    Article  Google Scholar 

  6. Masmoudi H, Coirault C, Demoule A, Mayaux J, Beuvin M, Romero N, Assouad J, Similowski T. Can phrenic stimulation protect the diaphragm from mechanical ventilation-induced damage? Eur Respir J. 2013;42:280–3.

    Article  Google Scholar 

  7. Yang MR, Wang HT, Han GW, Chen LH, Huang LN, Jiang JH, Li ST. Phrenic nerve stimulation protects against mechanical ventilation-induced diaphragm dysfunction in rats. Muscle Nerve. 2013;48:958–62.

    Article  CAS  Google Scholar 

  8. DiMarco AF, Onders RP, Ignagni A, Kowalski KE, Mortimer JT. Phrenic nerve pacing via intramuscular diaphragm electrodes in tetraplegic subjects. Chest. 2005;127:671–8.

    Article  Google Scholar 

  9. Evans D, Shure D, Clark L, Criner GJ, Dres M, de Abreu MG, Laghi F, McDonagh D, Petrof B, Nelson T, Similowski T. Temporary transvenous diaphragm pacing vs. standard of care for weaning from mechanical ventilation: study protocol for a randomized trial. Trials. 2019;20:60.

    Article  Google Scholar 

  10. Dres M, Gama de Abreu M, Merdji H, Muller-Redetzky H, Dellweg D, Randerath WJ, Mortaza S, Jung B, Bruells C, Morer O, et al. Randomised clinical study of temporary transvenous phrenic nerve stimulation in difficult-to-Wean patients. Am J Respir Crit Care Med. 2022;205(10):1169–78.

    Article  Google Scholar 

  11. Similowski T, Fleury B, Launois S, Cathala HP, Bouche P, Derenne JP. Cervical magnetic stimulation: a new painless method for bilateral phrenic nerve stimulation in conscious humans. J Appl Physiol. 1985;1989(67):1311–8.

    Google Scholar 

  12. Mills GH, Kyroussis D, Hamnegard CH, Polkey MI, Green M, Moxham J. Bilateral magnetic stimulation of the phrenic nerves from an anterolateral approach. Am J Respir Crit Care Med. 1996;154:1099–105.

    Article  CAS  Google Scholar 

  13. Polkey MI, Duguet A, Luo Y, Hughes PD, Hart N, Hamnegard CH, Green M, Similowski T, Moxham J. Anterior magnetic phrenic nerve stimulation: laboratory and clinical evaluation. Intensive Care Med. 2000;26:1065–75.

    Article  CAS  Google Scholar 

  14. Sander BH, Dieck T, Homrighausen F, Tschan CA, Steffens J, Raymondos K. Electromagnetic ventilation: first evalution of a new method for artificial ventilation in humans. Muscle Nerve. 2010;42:305–10.

    Article  Google Scholar 

  15. Adler D, Gottfried SB, Bautin N, Mirkovic T, Schmidt M, Raux M, Pavlovic D, Similowski T, Demoule A. Repetitive magnetic stimulation of the phrenic nerves for diaphragm conditioning: a normative study of feasibility and optimal settings. Appl Physiol Nutr Metab. 2011;36:1001–8.

    Article  Google Scholar 

  16. Kuhn KF, Grunow JJ, Leimer P, Lorenz M, Berger D, Schefold JC, Weber-Carstens S, Schaller SJ. Assessment of magnetic flux density properties of electromagnetic noninvasive phrenic nerve stimulations for environmental safety in an ICU environment. Sci Rep. 2021;11:16317.

    Article  CAS  Google Scholar 

  17. Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, Crapo R, Enright P, van der Grinten CP, Gustafsson P, et al. Standardisation of spirometry. Eur Respir J. 2005;26:319–38.

    Article  CAS  Google Scholar 

  18. Quanjer PH, Stanojevic S, Cole TJ, Baur X, Hall GL, Culver BH, Enright PL, Hankinson JL, Ip MS, Zheng J, et al. Multi-ethnic reference values for spirometry for the 3–95-yr age range: the global lung function 2012 equations. Eur Respir J. 2012;40:1324–43.

    Article  Google Scholar 

  19. Wilson SH, Cooke NT, Edwards RH, Spiro SG. Predicted normal values for maximal respiratory pressures in caucasian adults and children. Thorax. 1984;39:535–8.

    Article  CAS  Google Scholar 

  20. Fan E, Del Sorbo L, Goligher EC, Hodgson CL, Munshi L, Walkey AJ, Adhikari NKJ, Amato MBP, Branson R, Brower RG, et al. An Official American Thoracic Society/European Society of Intensive Care Medicine/Society of Critical Care medicine clinical practice guideline: mechanical ventilation in adult patients with acute respiratory distress syndrome. Am J Respir Crit Care Med. 2017;195:1253–63.

    Article  Google Scholar 

  21. Similowski T, Straus C, Attali V, Duguet A, Derenne JP. Cervical magnetic stimulation as a method to discriminate between diaphragm and rib cage muscle fatigue. J Appl Physiol. 1985;1998(84):1692–700.

    Google Scholar 

  22. Sotak M, Roubik K, Henlin T, Tyll T. Phrenic nerve stimulation prevents diaphragm atrophy in patients with respiratory failure on mechanical ventilation. BMC Pulm Med. 2021;21:314.

    Article  Google Scholar 

  23. Verin E, Series F, Locher C, Straus C, Zelter M, Derenne JP, Similowski T. Effects of neck flexion and mouth opening on inspiratory flow dynamics in awake humans. J Appl Physiol. 1985;2002(92):84–92.

    Google Scholar 

  24. Bellemare F, Grassino A. Effect of pressure and timing of contraction on human diaphragm fatigue. J Appl Physiol Respir Environ Exerc Physiol. 1982;53:1190–5.

    CAS  Google Scholar 

Download references

Acknowledgements

We would like to acknowledge the time, effort and tolerance of our participants during this difficult period of the COVID-19-pandemic, Prof. Dominique De Quervain (University of Basel, Switzerland) for providing the Brainsight Unit, Dr. Raniero Pittini and his team (Switzerland Innovation Park Biel/Bienne) for construction of the coil-fixation unit and providing Brainsight feedback instructions, Prof. Thomas Niederhauser and his team (Bern University of Applied Sciences, Switzerland) for valuable discussions and support, Daniel Leemann (Neurolite AG, Switzerland) for technical support, STIMIT AG (Switzerland) for technical, logistical and financial support, as well as Innosuisse (34221.1 IP-LS) for financial support.

Funding

Open access funding provided by Swiss Federal Institute of Technology Zurich Innosuisse (34221.1 IP-LS) provided financial support for the project.

Author information

Authors and Affiliations

Authors

Contributions

KGPJMB, PAE and CMS conceived the study. KGPJMB, PAE and PS performed the data acquisition. KGPJMB, PAE and CMS are responsible for data analysis and interpretation. KGPJMB drafted the manuscript, KGPJMB, PAE, and CMS revised the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Christina M. Spengler.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the Cantonal Ethics Committee of Zurich (Project ID 2019-01990) and registered on clinicaltrials.gov (NCT04176744). All participants signed an informed consent form and the study conformed with the Declaration of Helsinki.

Consent for publication

All participants in Fig. 2 and both online supplemental videos (Additional files 1 and 2) of this manuscript provided written informed consent to use their image.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Video S1. Three spontaneous breaths and a single RMSBAMPS.

Additional file 2: Video S2. RMSBAMPS ventilation.

Additional file 3.

Correlations between tidal volumes in response to single-train rapid bilateral anterior magnetic phrenic nerve stimulation and anthropometric variables.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Boyle, K.G.P.J.M., Eichenberger, P.A., Schön, P. et al. Inspiratory response and side-effects to rapid bilateral magnetic phrenic nerve stimulation using differently shaped coils: implications for stimulation-assisted mechanical ventilation. Respir Res 23, 357 (2022). https://doi.org/10.1186/s12931-022-02251-y