Skip to main content

Table 3 Prevalence of upper airway collapse and dental pain during single-train rapid bilateral anterior magnetic phrenic nerve stimulation (RMSBAMPS)

From: Inspiratory response and side-effects to rapid bilateral magnetic phrenic nerve stimulation using differently shaped coils: implications for stimulation-assisted mechanical ventilation

 

Upper airway collapse

Dental pain

 

Occurrences of upper airway collapse

Total number of RMSBAMPS trains conducted

Relative occurrence (%)

Occurrences of dental pain

Total number of dental pain ratings

Relative occurrence (%)

Stimulation-frequency

 15 Hz

71

237

30

22

106

21

 20 Hz

43

218

20

20

94

21

 25 Hz

25

188

13

22

86

26

 30 Hz

36

185

19

19

81

23

Stimulator-output

 20%

86

321

27

6

140

4

 30%

62

274

23

25

122

20

 40%

19

158

12

25

71

35

 50%

7

56

13

19

26

73

 60%

1

17

6

7

7

100

 70%

0

2

0

1

1

100

Sum for each coil

 Parabolic

49

214

23

19

94

20

 D-shape

78

280

28

12

123

10

 Butterfly

48

334

14

52

150

35

  1. Hz: Hertz. Values are absolute counts or percentages, as specified