Monitoring of noninvasive ventilation: comparative analysis of different strategies

Abstract

Background

Noninvasive ventilation (NIV) represents an effective treatment for chronic respiratory failure. However, empirically determined NIV settings may not achieve optimal ventilatory support. Therefore, the efficacy of NIV should be systematically monitored. The minimal recommended monitoring strategy includes clinical assessment, arterial blood gases (ABG) and nocturnal transcutaneous pulsed oxygen saturation (SpO2). Polysomnography is a theoretical gold standard but is not routinely available in many centers. Simple tools such as transcutaneous capnography (TcPCO2) or ventilator built-in software provide reliable informations but their role in NIV monitoring has yet to be defined. The aim of our work was to compare the accuracy of different combinations of tests to assess NIV efficacy.

Methods

This retrospective comparative study evaluated the efficacy of NIV in consecutive patients through four strategies (A, B, C and D) using four different tools in various combinations. These tools included morning ABG, nocturnal SpO2, TcPCO2 and data provided by built-in software via a dedicated module. Strategy A (ABG + nocturnal SpO2), B (nocturnal SpO2 + TcPCO2) and C (TcPCO2 + builtin software) were compared to strategy D, which combined all four tools (NIV was appropriate if all four tools were normal).

Results

NIV was appropriate in only 29 of the 100 included patients. Strategy A considered 53 patients as appropriately ventilated. Strategy B considered 48 patients as appropriately ventilated. Strategy C misclassified only 6 patients with daytime hypercapnia.

Conclusion

Monitoring ABG and nocturnal SpO2 is not enough to assess NIV efficacy. Combining data from ventilator built-in software and TcPCO2 seems to represent the best strategy to detect poor NIV efficacy.

Trial registration Institutional Review Board of the Société de Pneumologie de Langue Française (CEPRO 2016 Georges)

Background

Non-invasive ventilation (NIV) is recognized as an effective treatment of chronic hypercapnic respiratory failure (CHRF) [1]. Due to growing evidence of NIV efficacy in a broad range of indications as well as increasing availability of high performance and user-friendly home ventilators, the number of patients receiving NIV at home has been regularly increasing over the past 30 years [2,3,4]. When NIV is initiated to treat CHRF, ventilator settings are empirically determined based on the underlying disease, patient tolerance and diurnal changes in arterial blood gases (ABG) [5]. However, NIV is usually applied during the night. As a result, daytime adjustment of ventilator settings may not achieve optimal nocturnal ventilatory support. This can be explained by sleep-related changes in breathing. Sleep induces modifications in ventilatory control, respiratory muscle recruitment and upper airway patency, which may all affect ventilatory function especially in patients with CHRF [6]. Moreover, applying intermittent positive pressure may by itself trigger abnormal respiratory events [7]. For instance, reduction of ventilatory drive with or without glottic closure, residual upper airway obstruction and patient-ventilator asynchrony can all compromise the efficacy of NIV [7]. Furthermore, as NIV uses a non-airtight system, unintentional leaks are frequent [8]. Leaks during NIV can interfere with patient-ventilator interaction [9]. These respiratory events are frequent under NIV [8, 10,11,12,13] and may have an impact on prognosis [14,15,16].

Therefore, NIV should be systematically monitored. However, optimal modalities for monitoring of long-term ventilated patients remain a matter of debate. Hence, physicians may adopt different approaches to assess NIV performance. Some authors suggest that complete polysomnography (PSG) under NIV should be performed for each patient under NIV to verify its efficacy [7, 17]. This technique is not feasible in many centres on a routine basis. In contrast, the 2010 American Academy Sleep Medicine (AASM) recommendations for best clinical practices state that patients on long term NIV should be assessed regularly with measures of oxygenation and ventilation (i.e.: ABG, nocturnal pulse oximetry, end tidal CO2 or transcutaneous capnography) [18, 19]. Over the past years, the use of TcPCO2 has been simplified. Home ventilators have built-in software that provide detailed information on relevant ventilator parameters to assess the efficacy of NIV. A step-by-step strategy starting by ABG and nocturnal SpO2 has been proposed by the SomnoNIV group [19]. However, few studies have evaluated these proposed monitoring strategies in clinical practice [20].

This study aimed to compare the accuracy of four different strategies using four easily available assessment tools in different combinations to determine NIV efficacy during elective evaluations of patients on long-term NIV.

Methods

All patients under long-term home NIV followed by the Pulmonary Department of Dijon University Hospital are hospitalized electively for one night on a regular basis to assess efficacy of their NIV. These admissions are scheduled by the attending specialist every 3 to 12 months: intervals depend on the underlying respiratory disease and its progression rate, prior assessment of NIV efficacy or tolerance and intercurrent medical events.

In this retrospective comparative study, we included consecutive patients treated with long term NIV and hospitalized in our unit for an elective follow-up visit over a 1 year period. Inclusion criteria were: use of a home bi-level pressure support ventilator (VPAP™, ResMed, North Ryde, Australia) and being in a stable clinical condition for at least 3 months prior to inclusion.

Exclusion criteria included: age below 18 years, oxygen supplementation, use of a ventilator from other manufacturers, mean daily NIV use of less than 4 h per night, inability to cooperate and change in NIV treatment in the preceding 3 months.

NIV was evaluated with usual ventilator settings and interface. We simultaneously recorded overnight for each patient four monitoring tools: (1) morning ABG measured during spontaneous breathing by puncture of the radial artery during the first hour after disconnection from the ventilator, (2) nocturnal pulsed oxygen saturation (SpO2; Nonin model 8500 oximeter, Nonin Medical, Plymouth, MN, USA), (3) transcutaneous capnography (TcPCO2: Tosca®, Radiometer, Copenhagen, Denmark) and (4) data from a simplified monitoring module coupled to their portable ventilator (Reslink™, ResMed). Data from the ventilator software were collected on a Smart Media card (Scandisk, Milpita, CA, USA) then downloaded with Rescan™ software (ResMed, North Ryde, Australia). The software provided an accurate estimation of non-intentional air leaks (i.e. leaks exceeding what was expected from the exhalation valve of the interface used) [8]. The additional connection of a pulse oximeter allowed simultaneous recording of nocturnal SpO2.

Thresholds used to interpret results of the four monitoring tools were the following: (1) ABG: PaCO2 ≥ 45 mmHg; (2) nocturnal SpO2: time spent with SpO2 < 90% for ≥ 30% of the total recording time [21]; (3) transcutaneous capnography: mean TcPCO2 ≥ 50 mmHg [22, 23] and (4) data from built-in ventilator software: leaks (> 24 l/min for > 20% of total recording time), continuous desaturation (SpO2 < 90% for > 30% of the recording) and cumulated desaturation dips (> 3% during > 10% of the trace) [8].

We evaluated the efficacy of NIV through four strategies (A, B, C and D) using the results of four different tools, in different combinations: strategy A combined ABG and nocturnal SpO2, the minimal recommended monitoring combination [19]; strategy B combined nocturnal SpO2 and TcPCO2: since transcutaneous capnography provides SpO2 and TcPCO2 simultaneously, both parameters could be analyzed concurrently; strategy C combined TcPCO2 and data from built-in ventilator software and strategy D associated all the available tools (i.e. ABG, nocturnal SpO2, TcPCO2 and data from ventilator software). Strategy D is used to classify patients as appropriately ventilated or not. If none of the above-mentioned criteria were fulfilled, NIV was considered effective.

The St. Mary’s Hospital questionnaire was completed in the morning after the overnight assessment to evaluate subjective sleep quality on a 12 point scale [24]. Another questionnaire assessed the self-perceived quality of ventilation using an eight-item visual analogic scale (10 points per item) covering three domains: patient-ventilator synchronisation, efficacy and leaks [25]. Higher values indicated better treatment comfort, with a maximum score of 80.

The study was approved by the Institutional Review Board of the Société de Pneumologie de Langue Française.

Statistical analysis

Statistical analyses were performed using SigmaPlot 13 software (Systat Software, San Jose, CA, USA). The normality of the distribution of the variables analysed was assessed using the Kolmogorov–Smirnov test. As most data were not normally distributed, we reported results as median and quartiles and used non-parametric tests. We used the Mann Whitney’s U test to compare “appropriately” and “inappropriately” ventilated patients for continuous variables. Categorial variables (gender, interfaces) were compared using a χ2 test. For comparisons between three or more groups (classification of patients according to the aetiology of chronic respiratory failure), we used the Kruskal–Wallis test; subsequent paired comparisons were made using a post-hoc Dunn’s analysis. Statistical significance was set at p < 0.05 or p < 1 − (1 − α)1/k for multiple comparisons where α = 0.05 and k denotes the number of comparisons.

The agreement between different methods of NIV monitoring and the strategy D was evaluated with Cohen’s kappa coefficient [26].

We used receiver operating characteristic (ROC) curves to evaluate the performance of nocturnal SpO2 and ABG to identify patients classified as adequately ventilated according to strategy D. We considered agreement to be sufficient if the lower bound of 95% confidence interval for the area under the ROC curve was > 0.7. ROC curve analyses were also used to determine the most suitable threshold values of mean nocturnal SpO2 and morning PaCO2 for assessing NIV efficacy.

Results

One hundred and thirty-four patients were screened. Two subjects were excluded due to corruption of raw data from the ventilator software. Thirty-two patients under oxygen therapy were also excluded from further analyses. These subjects suffered more often from obstructive lung diseases (OLD) and presented more severe diurnal and nocturnal hypercapnia (p < 0.001).

Study population

The remaining 100 patients were treated with NIV for OLD (n = 25), chest wall diseases (CWD, n = 29) and neuromuscular diseases (NMD, n = 46) according to the Eurovent diagnostic groups [2] (Table 1). Demographic characteristics, ABG, TcPCO2 and ventilator settings are summarized in Table 2. As expected, NMD patients were younger, had a lower BMI and required lower levels of pressure support to reach more effective control of diurnal and nocturnal hypercapnia. Nasal masks were used more frequently in this group than in OLD or CWD subjects (p < 0.05).

Table 1 Characteristics of the studied population: indications for noninvasive ventilation according to Eurovent categories
Table 2 Characteristics of the studied population: demographic data, diurnal and nocturnal gas exchanges and ventilator settings

Assessment of NIV efficacy

TcPCO2 revealed significant nocturnal hypoventilation in 27% of the patients. Among them, 6% had normal ABG and 12% had normal nocturnal SpO2. Data from built-in ventilator software were abnormal in 57% of the patients. Leaks represented the most common abnormality (28%).

Table 3 compares the performances of different strategies. NIV was appropriate in only 29% of patients. No significant differences were found regarding ventilator settings or interfaces between appropriately and inappropriately ventilated patients. NIV compliance did not differ significantly between appropriately and inappropriately ventilated patients (8.5 [6.9–10] vs. 7.5 [6.1–9.9] hours per night, respectively).

Table 3 Proportion of patients considered as appropriately ventilated according to tests used alone or in various strategies

With strategy A, 53% of patients were considered appropriately ventilated. Among 48% of patients with normal results using strategy B, data from built-in ventilator software identified major leaks in 18% and significant drops in SpO2 associated with decreases in flow despite effective ventilator pressure in 10% of patients.

When using strategy C, NIV was considered appropriate in 35% of patients. Among them, only 6% had abnormal ABG and were misclassified. Strategy C performed better than strategies A or B for classifying appropriately vs. inappropriately ventilated patients (Cohen’s kappa coefficient, к for strategy A vs. D: 0.56 [0.41–0.71]; strategy B vs. D: 0.601 [0.436–0.755]; strategy C vs. D: 0.94 [0.86–1]).

Optimal threshold values for PaCO 2 and SpO 2 for identifying suboptimal NIV according to strategy D

Table 4 presents ROC curve analysis of optimal threshold value of ABG and nocturnal SpO2 for identifying appropriately ventilated patients (defined by strategy D).

Table 4 Results of receiver operating characteristic curve analyses for mean nocturnal SpO2, time spent with SpO2 < 90% and morning PaCO2 for the detection of inappropriate NIV (according to strategy D)

A morning PaCO2 value of 42 mmHg was the best threshold for identifying appropriate NIV (Fig. 1a): 69% of the patients were correctly classified using this value.

Fig. 1
figure 1

ROC curve of morning PaCO2 (a) and time spent with SpO2 below 90% (b) predicting NIV efficacy established by strategy D

The best threshold for time spent with SpO2 below 90% was 5% (Fig. 1b): 63% of the patients were correctly classified using this value. Higher values for time spent with SpO2 below 90% had a lower sensitivity with a similar specificity.

Subjective assessment of quality of sleep and comfort of ventilation

Perceive quality of sleep (Fig. 2a) and comfort of ventilation (Fig. 2b) did not differ significantly between appropriately and inappropriately ventilated patients. Neuromuscular patients reported a worse quality of sleep and increased fragmentation (see Additional file 1 for perceived sleep quality and comfort of ventilation according to Eurovent categories).

Fig. 2
figure 2

Patient’s rating of quality of sleep and ventilation assessed by St. Mary’s Hospital Questionnaire (a) and eight visual analogic scales (b) according to objective efficacy of NIV

Discussion

In this real-life study, we compared different strategies to assess the efficacy of NIV. Our results suggest that using a combination of daytime ABG and nocturnal SpO2 (referred to as strategy A, proposed by the group of experts [19]) was not sensitive enough to assess NIV efficacy. A significant part of this group had residual nocturnal abnormalities under NIV (hypoventilation, unintentional leaks or abnormal events). In this group, withholding from performing further NIV testing could be deleterious. A combination of TcPCO2 and data from ventilator software, referred to as strategy C, was the most accurate non-invasive strategy for assessing NIV efficacy.

Improving NIV efficacy is an important issue in patients with long-term NIV: residual respiratory events under NIV may have a negative impact on patient-related outcomes such as symptoms, health-related quality of life and survival. Nocturnal hypoventilation is associated with a decreased survival rate, especially in neuromuscular diseases [14, 16], as well as adverse neuro-cognitive and cardiovascular consequences in chronic respiratory failure [27]. Leaks above 0.4 l/s [28] may induce patient-ventilator asynchrony [12, 29], alter quality of sleep [30,31,32,33] and potentially decrease health-related quality of life. Abnormal respiratory events under NIV (upper airway obstructive events with or without nocturnal desaturations or residual hypoventilation or symptoms) are associated with a decreased survival rate in patients suffering from amyotrophic lateral sclerosis (ALS) [15].

To detect residual nocturnal hypoventilation, we suggest using TcPCO2 instead of morning ABG. In ventilated patients, PaCO2 measured by arterial puncture may not provide an accurate picture of the overnight time course of PaCO2 [19, 22]. Several studies have shown that continuous TcPCO2 recording is well correlated with arterial measurements in chronic respiratory failure under NIV [10, 34, 35].

Experts propose different thresholds to assess the efficacy of NIV but little evidence substantiates the relevance of these values. Regarding TcPCO2, several thresholds have been suggested to define significant nocturnal hypercapnia: maximal TcPCO2 > 49 mmHg [36, 37]; TcPCO2 > 49 mmHg for > 10% of recording time [22]; TcPCO2 > 55 mmHg for ≥ 10 min or an increase in TcPCO2 ≥ 10 mmHg above awake supine value to a value exceeding 50 mmHg for ≥ 10 min [18]. Clinically relevant threshold values may differ according to 1/the method and device used, 2/the etiology of chronic respiratory failure, 3/the goal of TcPCO2 recording (i.e. to decide when NIV should be initiated or to monitor NIV efficacy) and 4/PCO2 levels when NIV is started. For example, prognosis is improved in COPD if NIV effectively reduces PaCO2 by more than 20% [38]. The thresholds used may also depend on the type of capnograph as bias between arterial and transcutaneous values changes according to the device used [39]. The device used in our study slightly overestimated PaCO2. The maximal bias published with this device was 5.6 ± 3 mmHg [40]. We therefore considered residual nocturnal hypoventilation as significant when mean TcPCO2 was ≥ 50 mmHg [41].

The clinical contribution of nocturnal transcutaneous capnography can be improved by simultaneously recording SpO2 [19]. Sampling rate and averaging of SpO2 and TcPCO2 recordings are different: SpO2 can detect short desaturations linked to short ventilatory events while TcPCO2 has a longer lag time but is an accurate tool to evaluate overnight trends in ventilation. Hence, both tools are complementary and devices used in clinical practice combine TcPCO2 and SpO2 sensors. However, capnography does not provide information about the underlying pathophysiological mechanisms. Furthermore, in a quarter of patients with normal TcPCO2 and SpO2 (strategy B), we found significant leaks or abnormal residual respiratory events (ie, flow reduction or patient-ventilator asynchronies). Our study confirms the additional contribution of data from ventilator software for the detection of these events. The accuracy of the ResScan™ system used to assess leaks has been confirmed in a bench model by our group and others [8, 42].

Our results suggest that using more severe thresholds for PaCO2 and NPO may compensate their lack of sensitivity. For instance, using a PaCO2 threshold value of 42 mmHg could increase the accuracy of ABG for the detection of nocturnal hypoventilation.

Time spent with a SpO2 below 90% is the most frequently used parameter to interpret nocturnal pulse oximetry, but threshold values vary considerably between authors and aetiologies. In non-ventilated patients suffering from chronic obstructive pulmonary disease (COPD), Levi Valensi et al. [43] documented a shorter survival in patients spending more than 30% of total sleep time with an SpO2 below 90%. More recently, Gonzalez-Bermejo et al. [14] showed that ALS patients under NIV had a better survival if less than 5% of NPO time was spent with an SpO2 < 90%. In our study, using a threshold of 5% increased the accuracy of NPO in detecting residual nocturnal hypoventilation.

An analysis combining the signals provided by TcPCO2 and data from ventilator software may be an interesting option for monitoring NIV, offering a noninvasive global estimation of NIV efficacy without requiring ABG. Moreover, this approach enables unattended assessment both at the hospital and at home without complex logistics. Failure to retrieve data is rare [44] and instrumental drift of TcPCO2 is a minor problem when used by an experienced team [20, 39, 45, 46]. Interpretation of the results is simple and further analysis of detailed raw data provided by ventilator software can help clarify the underlying mechanism implicated in NIV inefficacy. This may allow optimization of ventilator settings limiting PSG to more complex cases. Unfortunately, use of TcPCO2 is at present still limited by the cost of the devices.

We acknowledge a few limitations to our study. Firstly, we did not perform full PSG under NIV. Even if PSG allows the evaluation of patient-ventilator interactions and characterization of abnormal respiratory events occurring under NIV [7], the impact of these events on morbidity and related therapeutic end points remains speculative [47]. Furthermore, it does not provide an accurate estimation of alveolar ventilation per se, which is the main goal of ventilator assistance. It is also probable that leaks could be underscored by PSG.

Secondly, we excluded 32 patients with nocturnal NIV and oxygen therapy. Supplemental oxygen impacts on SpO2 values and reduces the amplitude of desaturations, decreasing the reliability of NPO to assess NIV efficacy. It must be noted that the majority of excluded patients suffered from chronic obstructive pulmonary disease.

Thirdly, NIV is considered beneficial if used more than 4 h per night (for ALS [48]; for COPD [49]; for obesity-hypoventilation syndrome [50]). We also excluded patients using NIV for less than 4 h per night. Poor compliance to NIV may result from discomfort related to leaks or a low perceived benefit of treatment. This could have underestimated the proportion of inadequately ventilated patients even if leaks represent the most frequent abnormality in our study.

Fourthly, we failed to show an impact of NIV efficacy on sleep quality or patient symptoms. Both scores employed for assessing comfort and quality of sleep have been previously used to assess subjective impact of changes in ventilator modes (volume-targeted versus conventional bi-level pressure support) [25]. Our results suggest that subjective assessment does not suffice for the detection of inappropriate ventilation. The poor correlation between residual respiratory events and patients’ perception has been previously reported [9, 10]. Finally, the impact of NIV efficacy on survival could not be assessed due to the heterogeneity of our population consisting of subgroups (OLD, CWD, NMD) with different prognoses. Further investigations are needed to identify which of the selected tools included significantly impacts on patient-related outcomes such as symptoms, health-related quality of life or survival.

In summary, this study shows that combining morning ABG and nocturnal SpO2 is not sufficient to accurately assess NIV efficacy. An alternative strategy combining data from ventilator software and TcPCO2 performed better for detecting inappropriate NIV without requiring ABG. Models of care for chronically ill patients living at home are evolving with tele-monitoring. TcPCO2 and ventilator software data are increasingly available at home. Moreover, their easy interpretation makes it feasible in real life and in a variety of clinical settings. This combination may be very useful in future strategies for long-term NIV monitoring.

Availability of supporting data

The datasets generated and analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ALS:

Amyotrophic lateral sclerosis

ABG:

Arterial blood gases

BMI:

Body mass index

CHRF:

Chronic hypercapnic respiratory failure

COPD:

Chronic obstructive pulmonary disease

CWD:

Chest wall diseases

NIV:

Noninvasive ventilation

NMD:

Neuromuscular disease

OLD:

Obstructive lung diseases

PaO2 :

Arterial dioxygen partial pressure

PaCO2 :

Arterial carbon dioxide partial pressure

PSG:

Polysomnography

SpO2 :

Transcutaneous pulsed oxygen saturation

TcPCO2 :

Transcutaneous carbon dioxide partial pressure

References

  1. Turkington PM, Elliott MW. Rationale for the use of non-invasive ventilation in chronic ventilatory failure. Thorax. 2000;55(5):417–23.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Lloyd-Owen SJ, Donaldson GC, Ambrosino N, Escarabill J, Farre R, Fauroux B, Robert D, Schoenhofer B, Simonds AK, Wedzicha JA. Patterns of home mechanical ventilation use in Europe: results from the Eurovent survey. Eur Respir J. 2005;25(6):1025–31.

    Article  CAS  PubMed  Google Scholar 

  3. Rabec C, Gonzalez-Bermejo J, Cuvelier A, Cervantes P, Foret D, Mounier L, Melloni B, Muir JF. Cohort of patients initiated to home ventilation. Observational and prospective study. Rev Mal Respir. 2018;35(1):88–93.

    Article  CAS  PubMed  Google Scholar 

  4. Cantero C, Adler D, Pasquina P, Uldry C, Egger B, Prella M, Younossian AB, Soccal PM, Pepin JL, Janssens JP. Long-term noninvasive ventilation in the Geneva Lake Area: indications, prevalence, and modalities. Chest. 2020;158(1):279–91.

    Article  PubMed  Google Scholar 

  5. Elliott MW. Non-invasive ventilation during sleep: time to define new tools in the systematic evaluation of the technique. Thorax. 2011;66(1):82–4.

    Article  PubMed  Google Scholar 

  6. Adler D, Janssens JP. The pathophysiology of respiratory failure: control of breathing, respiratory load, and muscle capacity. Respiration. 2019;97(2):93–104.

    Article  PubMed  Google Scholar 

  7. Gonzalez-Bermejo J, Perrin C, Janssens JP, Pepin JL, Mroue G, Leger P, Langevin B, Rouault S, Rabec C, Rodenstein D, et al. Proposal for a systematic analysis of polygraphy or polysomnography for identifying and scoring abnormal events occurring during non-invasive ventilation. Thorax. 2012;67(6):546–52.

    Article  CAS  PubMed  Google Scholar 

  8. Rabec C, Georges M, Kabeya NK, Baudouin N, Massin F, Reybet-Degat O, Camus P. Evaluating noninvasive ventilation using a monitoring system coupled to a ventilator: a bench-to-bedside study. Eur Respir J. 2009;34(4):902–13.

    Article  CAS  PubMed  Google Scholar 

  9. Ramsay M. Patient ventilator asynchrony and sleep disruption during non-invasive ventilation. J Thorac Dis. 2018;10(Suppl 1):S80–5.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Aarrestad S, Qvarfort M, Kleiven AL, Tollefsen E, Skjonsberg OH, Janssens JP. Sleep related respiratory events during non-invasive ventilation of patients with chronic hypoventilation. Respir Med. 2017;132:210–6.

    Article  PubMed  Google Scholar 

  11. Guo YF, Sforza E, Janssens JP. Respiratory patterns during sleep in obesity-hypoventilation patients treated with nocturnal pressure support: a preliminary report. Chest. 2007;131(4):1090–9.

    Article  PubMed  Google Scholar 

  12. Caldarelli V, Borel JC, Khirani S, Ramirez A, Cutrera R, Pepin JL, Fauroux B. Polygraphic respiratory events during sleep with noninvasive ventilation in children: description, prevalence, and clinical consequences. Intensive Care Med. 2013;39(4):739–46.

    Article  PubMed  Google Scholar 

  13. Ramsay M, Mandal S, Suh ES, Steier J, Douiri A, Murphy PB, Polkey M, Simonds A, Hart N. Parasternal electromyography to determine the relationship between patient-ventilator asynchrony and nocturnal gas exchange during home mechanical ventilation set-up. Thorax. 2015;70(10):946–52.

    Article  PubMed  Google Scholar 

  14. Gonzalez-Bermejo J, Morelot-Panzini C, Arnol N, Meininger V, Kraoua S, Salachas F, Similowski T. Prognostic value of efficiently correcting nocturnal desaturations after one month of non-invasive ventilation in amyotrophic lateral sclerosis: a retrospective monocentre observational cohort study. Amyotroph Lateral Scler Frontotemporal Degener. 2013;14(5–6):373–9.

    Article  PubMed  Google Scholar 

  15. Georges M, Attali V, Golmard JL, Morelot-Panzini C, Crevier-Buchman L, Collet JM, Tintignac A, Morawiec E, Trosini-Desert V, Salachas F, et al. Reduced survival in patients with ALS with upper airway obstructive events on non-invasive ventilation. J Neurol Neurosurg Psychiatry. 2016;87(10):1045–50.

    Article  PubMed  Google Scholar 

  16. Ogna A, Nardi J, Prigent H, Quera Salva MA, Chaffaut C, Lamothe L, Chevret S, Annane D, Orlikowski D, Lofaso F. Prognostic value of initial assessment of residual hypoventilation using nocturnal capnography in mechanically ventilated neuromuscular patients: a 5-year follow-up study. Front Med (Lausanne). 2016;3:40.

    Google Scholar 

  17. Gonzalez MM, Parreira VF, Rodenstein DO. Non-invasive ventilation and sleep. Sleep Med Rev. 2002;6(1):29–44.

    Article  PubMed  Google Scholar 

  18. Berry RB, Chediak A, Brown LK, Finder J, Gozal D, Iber C, Kushida CA, Morgenthaler T, Rowley JA, Davidson-Ward SL, et al. Best clinical practices for the sleep center adjustment of noninvasive positive pressure ventilation (NPPV) in stable chronic alveolar hypoventilation syndromes. J Clin Sleep Med. 2010;6(5):491–509.

    Article  PubMed  Google Scholar 

  19. Janssens JP, Borel JC, Pepin JL, Somno NIVG. Nocturnal monitoring of home non-invasive ventilation: the contribution of simple tools such as pulse oximetry, capnography, built-in ventilator software and autonomic markers of sleep fragmentation. Thorax. 2011;66(5):438–45.

    Article  PubMed  Google Scholar 

  20. Aarrestad S, Qvarfort M, Kleiven AL, Tollefsen E, Skjonsberg OH, Janssens JP. Diagnostic accuracy of simple tools in monitoring patients with chronic hypoventilation treated with non-invasive ventilation; a prospective cross-sectional study. Respir Med. 2018;144:30–5.

    Article  PubMed  Google Scholar 

  21. Levi-Valensi P, Weitzenblum E, Rida Z, Aubry P, Braghiroli A, Donner C, Aprill M, Zielinski J, Wurtemberger G. Sleep-related oxygen desaturation and daytime pulmonary haemodynamics in COPD patients. Eur Respir J. 1992;5(3):301–7.

    CAS  PubMed  Google Scholar 

  22. Paiva R, Krivec U, Aubertin G, Cohen E, Clement A, Fauroux B. Carbon dioxide monitoring during long-term noninvasive respiratory support in children. Intensive Care Med. 2009;35(6):1068–74.

    Article  PubMed  Google Scholar 

  23. Simonds AK. Home ventilation. Eur Respir J Suppl. 2003;47:38s–46s.

    Article  CAS  PubMed  Google Scholar 

  24. Ellis BW, Johns MW, Lancaster R, Raptopoulos P, Angelopoulos N, Priest RG. The St. Mary’s Hospital sleep questionnaire: a study of reliability. Sleep. 1981;4(1):93–7.

    Article  CAS  PubMed  Google Scholar 

  25. Janssens JP, Metzger M, Sforza E. Impact of volume targeting on efficacy of bi-level non-invasive ventilation and sleep in obesity-hypoventilation. Respir Med. 2009;103(2):165–72.

    Article  PubMed  Google Scholar 

  26. Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159–74.

    Article  CAS  PubMed  Google Scholar 

  27. Fletcher E, Donner C, Midgren B, Zielinski J, Levi-Valensi P, Braghiroli A, Rida Z, Miller C. Survival in COPD patients with a daytime PaO2 greater than 60 mm Hg with and without nocturnal oxyhemoglobin desaturation. CHEST J. 1992;101(3):649–55.

    Article  CAS  Google Scholar 

  28. Teschler H. Monitoring of the home mechanical ventilated patient. Eur Respir Monogr. 2001;6:274–80.

    Google Scholar 

  29. Carteaux G, Lyazidi A, Cordoba-Izquierdo A, Vignaux L, Jolliet P, Thille AW, Richard JC, Brochard L. Patient-ventilator asynchrony during noninvasive ventilation: a bench and clinical study. Chest. 2012;142(2):367–76.

    Article  PubMed  Google Scholar 

  30. Teschler H, Stampa J, Ragette R, Konietzko N, Berthon-Jones M. Effect of mouth leak on effectiveness of nasal bilevel ventilatory assistance and sleep architecture. Eur Respir J. 1999;14(6):1251–7.

    Article  CAS  PubMed  Google Scholar 

  31. Rabec C, Rodenstein D, Leger P, Rouault S, Perrin C, Gonzalez-Bermejo J. Ventilator modes and settings during non-invasive ventilation: effects on respiratory events and implications for their identification. Thorax. 2011;66:170–8.

    Article  PubMed  Google Scholar 

  32. Meyer TJ, Pressman MR, Benditt J, McCool FD, Millman RP, Natarajan R, Hill NS. Air leaking through the mouth during nocturnal nasal ventilation: effect on sleep quality. Sleep. 1997;20(7):561–9.

    Article  CAS  PubMed  Google Scholar 

  33. Pépin J, Borel J, Janssens J, Tamisier R, Levy P. Sleep and NIV: monitoring of the patient under home ventilation. Eur Respir Mon. 2008;41:350–66.

    Google Scholar 

  34. Maniscalco M, Zedda A, Faraone S, Carratu P, Sofia M. Evaluation of a transcutaneous carbon dioxide monitor in severe obesity. Intensive Care Med. 2008;34(7):1340–4.

    Article  PubMed  Google Scholar 

  35. Török SS, Leuppi D Jr, Baty F, Tamm M, Chhajed PN. Combined oximetry-cutaneous capnography in patients assessed for long-term oxygen therapy. CHEST J. 2008;133(6):1421–5.

    Article  Google Scholar 

  36. Ward S, Chatwin M, Heather S, Simonds AK. Randomised controlled trial of non-invasive ventilation (NIV) for nocturnal hypoventilation in neuromuscular and chest wall disease patients with daytime normocapnia. Thorax. 2005;60(12):1019–24.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  37. Nardi J, Prigent H, Adala A, Bohic M, Lebargy F, Quera-Salva MA, Orlikowski D, Lofaso F. Nocturnal oximetry and transcutaneous carbon dioxide in home-ventilated neuromuscular patients. Respir Care. 2012;57(9):1425–30.

    Article  PubMed  Google Scholar 

  38. Murphy PB, Rehal S, Arbane G, Bourke S, Calverley PMA, Crook AM, Dowson L, Duffy N, Gibson GJ, Hughes PD, et al. Effect of home noninvasive ventilation with oxygen therapy vs oxygen therapy alone on hospital readmission or death after an acute COPD exacerbation: a randomized clinical trial. JAMA. 2017;317(21):2177–86.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Aarrestad S, Tollefsen E, Kleiven AL, Qvarfort M, Janssens JP, Skjonsberg OH. Validity of transcutaneous PCO2 in monitoring chronic hypoventilation treated with non-invasive ventilation. Respir Med. 2016;112:112–8.

    Article  PubMed  Google Scholar 

  40. Stein N, Matz H, Schneeweiss A, Eckmann C, Roth-Isigkeit A, Huppe M, Gehring H. An evaluation of a transcutaneous and an end-tidal capnometer for noninvasive monitoring of spontaneously breathing patients. Respir Care. 2006;51(10):1162–6.

    PubMed  Google Scholar 

  41. Simonds AK. Chronic hypoventilation and its management. Eur Respir Rev. 2013;22(129):325–32.

    Article  PubMed  Google Scholar 

  42. Contal O, Vignaux L, Combescure C, Pepin JL, Jolliet P, Janssens JP. Monitoring of noninvasive ventilation by built-in software of home bilevel ventilators: a bench study. Chest. 2012;141(2):469–76.

    Article  PubMed  Google Scholar 

  43. Levi-Valensi P, Aubry P, Rida Z. Nocturnal hypoxemia and long-term oxygen therapy in COPD patients with daytime PaO2 60–70 mmHg. Lung. 1990;168(Suppl):770–5.

    Article  PubMed  Google Scholar 

  44. Bauman KA, Kurili A, Schmidt SL, Rodriguez GM, Chiodo AE, Sitrin RG. Home-based overnight transcutaneous capnography/pulse oximetry for diagnosing nocturnal hypoventilation associated with neuromuscular disorders. Arch Phys Med Rehabil. 2013;94(1):46–52.

    Article  PubMed  Google Scholar 

  45. Felemban O, Leroux K, Aubertin G, Miandy F, Damagnez F, Amorim B, Ramirez A, Fauroux B. Value of gas exchange recording at home in children receiving non-invasive ventilation. Pediatr Pulmonol. 2011;46(8):802–8.

    Article  PubMed  Google Scholar 

  46. Georges M, Nguyen-Baranoff D, Griffon L, Foignot C, Bonniaud P, Camus P, Pepin JL, Rabec C. Usefulness of transcutaneous PCO2 to assess nocturnal hypoventilation in restrictive lung disorders. Respirology. 2016;21(7):1300–6.

    Article  PubMed  Google Scholar 

  47. Hannan LM, Rautela L, Berlowitz DJ, McDonald CF, Cori JM, Sheers N, Chao C, O’Donoghue FJ, Howard ME. Randomised controlled trial of polysomnographic titration of noninvasive ventilation. Eur Respir J. 2019;53(5):1802118.

    Article  CAS  PubMed  Google Scholar 

  48. Bourke SC, Tomlinson M, Williams TL, Bullock RE, Shaw PJ, Gibson GJ. Effects of non-invasive ventilation on survival and quality of life in patients with amyotrophic lateral sclerosis: a randomised controlled trial. Lancet Neurol. 2006;5(2):140–7.

    Article  PubMed  Google Scholar 

  49. Borel JC, Pepin JL, Pison C, Vesin A, Gonzalez-Bermejo J, Court-Fortune I, Timsit JF. Long-term adherence with non-invasive ventilation improves prognosis in obese COPD patients. Respirology. 2014;19(6):857–65.

    Article  PubMed  Google Scholar 

  50. Mokhlesi B. Obesity hypoventilation syndrome: a state-of-the-art review. Respir Care. 2010;55(10):1347–62 (discussion 1363-1345).

    PubMed  Google Scholar 

Download references

Acknowledgements

None.

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

MG and CR designed the study. MG, CR, EM and GB collected data. MG, CR and SA analysed the results. MG took the lead in writing the manuscript. All authors discussed the results and contributed to the final version of the manuscript. CR, PB and JPJ provided critical feedback and helped shape the manuscript and design the figures. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Marjolaine Georges.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the Institutional Review Board of the Société de Pneumologie de Langue Française. Patients received a written information regarding this study and the use of their anonymized data for statistical purposes then they give their oral consent.

Consent for publication

No details relating to an individual person in the manuscript.

Competing interests

All the authors declare no competing interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

Patient’s rating quality of sleep and ventilation assessed by St. Mary’s Hospital Questionnaire and eight visual analogic scales according to aetiological groups and objective efficacy of NIV.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Georges, M., Rabec, C., Monin, E. et al. Monitoring of noninvasive ventilation: comparative analysis of different strategies. Respir Res 21, 324 (2020). https://doi.org/10.1186/s12931-020-01586-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12931-020-01586-8

Keywords