Micron-sized and submicron-sized aerosol deposition in a new ex vivo preclinical model

Abstract

Background

The knowledge of where particles deposit in the respiratory tract is crucial for understanding the health effects associated with inhaled drug particles.

Method

An ex vivo study was conducted to assess regional deposition patterns (thoracic vs. extrathoracic) of radioactive polydisperse aerosols with different size ranges [0.15 μm–0.5 μm], [0.25 μm–1 μm] and [1 μm–9 μm]. SPECT/CT analyses were performed complementary in order to assess more precisely the regional deposition of aerosols within the pulmonary tract.

Experiments were set using an original respiratory tract model composed of a human plastinated head connected to an ex vivo porcine pulmonary tract. The model was ventilated by passive expansion, simulating pleural depressions. Aerosol was administered during nasal breathing.

Results

Planar scintigraphies allowed to calculate the deposited aerosol fractions for particles in the three size ranges from sub-micron to micron The deposited fractions obtained, for thoracic vs. extra-thoracic regions respectively, were 89 ± 4 % vs. 11 ± 4 % for [0.15 μm–0.5 μm], 78 ± 5 % vs. 22 ± 5 % for [0.25 μm–1 μm] and 35 ± 11 % vs.65 ± 11 % for [1 μm–9 μm].

Conclusion

Results obtained with this new ex vivo respiratory tract model are in good agreement with the in vivo data obtained in studies with baboons and humans.

Background

The effectiveness of aerosol drug delivery mainly depends on the amount of active agent penetrating the respiratory tract as well as on its regional distribution. This information is crucial for predicting the beneficial health effects associated with inhaled drug particles. Many inhalation studies were mainly focused on the assessment of the total deposited fraction (often calculated by comparison between inhaled and exhaled fractions) or used computational approaches, for instance, the recent Euler-Lagrangian approach of Kannan et al. [1]. In their study, high fidelity computational simulations were performed over several breathing cycles to get the regional deposition for different particle sizes and an algorithm was devised to account for the re-entry of particles during the exhalation phase. However some attempts were devoted to the quantification of regional particle deposition [24]. Nevertheless, in vivo experimental studies on the regional deposition of submicron-sized particles are still quite scarce.

In vivo experiments using radioactive aerosols are widely considered as the gold standard in order to assess the regional deposition of airborne particles. However, obtaining rapidly an extensive experimental dataset in human using radioactive aerosols is difficult and questionable due to ethical restrictions. Thus, in vivo laboratory animal studies are an attractive alternative. On the one hand, rodents, like rats or mice, that are most often used for such inhalation experiments [5, 6] are disparate from the human airways by size, bronchial divisions, anatomy of upper airways and respiratory physiology [7]. As an example, the respiratory rate at rest is around 80 breaths per minute for rats compared to 15 breaths per minute for humans. In these conditions, researchers have to be careful when translating results from in vivo experiments on rodents to humans even if they are helpful to understand deposition mechanisms. On the other hand, radioactive aerosol deposition studies have also been performed using pigs or non-human primates because of their anatomical proximity with human airways [811]. However, the main limits of these in vivo laboratory animal studies are ethical restrictions as well as cost and technical constraints (e.g., anesthesia that is often necessary to perform inhalation experiments, specialized laboratories, etc.). Furthermore, the breathing pattern, known to be an important factor for inhalation efficiency (Inspired/Expired ratio, frequency, obstruction etc.), is difficult to control in spontaneously breathing pigs or non-human primates.

As an alternative to in vivo inhalation experiments on humans and laboratory animals, ex vivo anatomical models such as cadaver heads [12] or in vitro nasal casts using prototyping techniques [13, 14] have been developed. These models are very accurate to assess aerosol drug deposition in nasal cavities such as maxillary sinuses [1519]. Nasal casts are morphologically close to the human anatomy and easier to handle compared to in vivo preclinical models. However, even if these replicas are useful for inhalation studies in nasal cavities, very few in vitro or ex vivo anatomical models mimic a human-like respiratory tract including both extrathoracic [ET] and thoracic [TH] regions and a physiological breathing pattern.

This article focuses on an original solution to bridge the existing gap on anatomical respiratory models to assess regional aerosol deposition. This study used an ex vivo human-like respiratory tract model [20]. This chimeric model is composed of a human plastinated head linked to an ex vivo porcine pulmonary tract ventilated artificially by inflation with a negative pressure inside a sealed enclosure (passive expiration). This cheaper and original approach without ethical restriction allows assessing regional aerosol deposition in both extrathoracic and pulmonary regions.

The clinical context of this study is the topical nasal drug delivery. When developing a drug product for nasal delivery, the aerodynamic features of the formulation have to ensure that drug particles are retained in the nasal cavity and not inhaled into the lung. For example, airborne drug delivery to the maxillary sinuses requires variable pressure application (known as ‘pulsating airflow’ aerosols) [1519, 21] but also small droplets (with a MMAD lower than 5 μm) [22]. The main challenge consists of delivering a high concentration of medication into the maxillary sinus cavities while at the same time preventing unwanted lung deposition observed with small particle delivery systems in order to minimize side effects. Therefore, the optimal droplet size for nasal administration should be based, among other things, on an accurate assessment of aerosol regional deposition in both extra thoracic and pulmonary regions.

The experiments were conducted using radioactive polydisperse aerosols in the [0.15 μm–0.5 μm], [0.25 μm–1 μm] and [1 μm–9 μm] ranges. We used exclusively a nasal administration of aerosol. We measured regional deposition in the TH region versus the ET region using planar scintigraphies and single-photon emission computed tomography (SPECT) [11, 13]. The secondary objective was to evaluate the reliability of this ex vivo model for aerosol deposition by comparison with known in vivo data in humans and baboons using the same polydisperse radioactive aerosols and similar nasal administration [11, 17].

Methods

Anatomical model

Experiments were performed using an ex vivo respiratory tract model. This model is composed of a human plastinated head linked to an ex vivo porcine pulmonary tract ventilating artificially by passive expansion with depressions. The laryngeal part is made of plastic tubes with a one-way valve simulating the resistance of vocal folds as inhalation is ongoing. An other one-way valve allows the path for exhalation as previously described [13]. The human plastinated head was anatomically, geometrically and aerodynamically characterized [12]. Physiological parameters obtained were consistent with the widely accepted human physiology [20]. Experimental conditions used were biomimetic with an original breathing technological process simulating the intrapleural depressions. Indeed, the chimeric model respired artificially by inflation with a negative pressure (passive expiration) applied inside an instrumented sealed enclosure developed specifically for this model [20]. The inhalation is due to the depression inside the instrumented sealed enclosure and the exhalation is passive by the return to atmospheric pressure by a pause of the depression generator. Moreover the duration of inhalation and exhalation in the respiratory cycle are chosen with shorter time duration for inhalation. In order to fit the human physiology, a ratio of 1.3 s for inhalation and 2.6 s for exhalation has been chosen, leading to a respiratory cycle of 4 s and consequently a respiratory rate of 15/min. In these conditions, the breathing parameters selected correspond to adult male physiology at rest: breathing rate of 15 breaths per minute, calculated mean tidal volume [VT] of 824 mL (SD 207 mL), inspiratory-to-expiratory time [I/E] ratio of 1/2, ET deadspace of 21 mL, tracheal internal diameter of 1.7 cm [2325]. The chimeric model satisfied the major functional and anatomical features necessary to study deposition of airborne particles [20]. The experimental setup for inhalation experiments is described in Fig. 1. The position of the exhalation filter is due to a particularity of the plastinated head (closed mouth) so obtained results correspond to an approximation of the exhaled fraction.

Fig. 1
figure 1

a Schematic of the experimental setup for nebulization using the human-like ex vivo preclinical model. 1) plastinated human head, 2) one-way valves, 3) plastic tubes, 4) expiratory filter, 5) porcine pulmonary tract, 6) plastic sealed enclosure, 7) respiratory pump, 8) nasal plug, 9) jet nebulizer, 10) compressor. b Photograph of the setup before the start of the aerosol experiments

Aerosol generation and characterization

Medical jet nebulizers are used to generate radioactive polydisperse aerosols from sub-micron to micron-sized particles. The nebulizing systems chosen are an Atomisor NL11 (DTF Medical, France), a modified Sidestream (Philips Respironics, Ref 12NEB400, England), and a Nanoneb (DTF Medical, France) operating at a flow rate of 6 L/min. The same experimental conditions concerning the generation of radioactive polydisperse aerosols were used as in previous in vivo studies performed in humans and baboons [11, 17].

Briefly, the nebulizer (Atomisor NL11, Sidestream or Nanoneb) was associated with a same AOHBOX air source compressor (Diffusion Technique Française, DTF Medical, Saint-Etienne, France). The nebulizers were connected to a nasal plug (C28 medium size, DTF Medical, France) usually employed in clinical practice. This plug ensures the interface connection between the nebulizer and plastinated head nostrils. For each experiment, nebulizers (n = 4) were loaded with 4 mL of a solution containing 740 MBq of 99mTc-DTPA (TechneScan DTPA, Mallinckrodt Medical, Petten, the Netherlands) and the duration of nebulization was standardized to 10 min as previously described [13, 18, 19, 21].

Radioactive aerosols are well characterized in terms of particle sizing and aerosol output according to protocols developed in previous works [11, 13]. Succinctly, particle size distributions of radioactive aerosols were determined using a gamma camera (Ecam; Siemens, Germany) coupled to a specific electrical low-pressure impactor (ELPI; Dekati, Finland) using the DLPI mode (without the corona charger of ELPI instrument). The DLPI mode allows avoiding possible biases induced by the corona charger (particle size selection phenomena and possible particle losses). The particles were impacted depending on their inertia-related aerodynamic diameter on one of the 12 size fraction stages of the impactor in the 30 nm - 10 μm range. Finally the radioactivity from aerosol particles deposited in each size-specific stage was quantified by scintigraphic imaging. It allowed direct tracing of the activity distribution as a function of aerosol aerodynamic size. Thus, the knowledge of the radioactivity dose deposited on each impactor stage permitted the calculation of:

  • the activity median aerodynamic diameter (AMAD) with geometric standard deviation (GSD),

  • the total dose of radioactivity delivered by the nebulizer (i.e. the nebulizer emitted fraction),

  • the fine particle (FP) fraction (including all particles with an aerodynamic diameter <2.5 μm) and the ultrafine particle (UFP) fraction (including all particles with an aerodynamic diameter <0.1 μm),

  • and the [d16–d84] particle size range (where d16 and d84 are respectively the particle diameters at the 16 and 84 % size cut-offs of the cumulative distribution).

Aerosol regional deposition

2D gamma-imaging acquisition and image analysis procedure

The planar scintigraphic images (matrix 256*256) were recorded with a variable angle dual detector SPECT/ CT (SYMBIA T2; Siemens, Knoxville, TN) equipped with a low-energy, high-resolution collimator (FWHM 8.3 mm at 10 cm); tested weekly for uniformity (UFOV 533 mm × 387 mm, CFOV 400 mm × 290 mm).

Before conducting the inhalation experiments, the initial radioactive dose filled in the nebulizer was quantified (scintigraphic images, 60-sec anterior/posterior, were acquired corresponding to the full and empty syringe). Once the inhalation experiments were performed, 180-sec anterior/posterior images of the experimental setup were acquired for each element: empty nebulizer, expiratory filter, human plastinated head and lungs as shown on Fig. 2.

Fig. 2
figure 2

Definition of the 2D imaging anatomical regions of interest (ROIs). Illustrative images of the planar scintigraphies obtained after each experiment. ROIs are delimited in red and the three green squares on each image correspond to the background evaluations

A region of interest (ROI) was delimited on the images with a correction of the background using the mean of three external ROIs. Calculations took into account the background radiation, physical decay of radioactivity and tissue attenuation correction factors.

Results were expressed in terms of the activity loaded into the nebulizer and the ratio between TH and ET regions corresponding respectively to the lungs with the tracheobronchial tree and to the head with the artificial larynx.

3D gamma-imaging acquisition and image analysis procedure

After acquiring 2D images with the same gamma camera (SYMBIA T2; Siemens, Knoxville, TN), SPECT and CT acquisitions were performed immediately (without moving the lungs) which have a 2-slice spiral CT for a more rapid and accurate attenuation correction and anatomical mapping. A 3D SPECT acquisition of the lungs was performed with 64 (2 × 32) projection images, each of 30s. Finally, a CT was performed with the following parameters: 130 kV, 90 mAs, 1.25 mm slice thickness, 0.9 mm increment, 1.6 mm pitch, and rotation time of 1.5 s. A multimodality computer platform (Symbia net; Siemens) was used for image reviews and manipulations. Both the transmission and emission scans were reconstructed using 3D OSEM by default (8 subsets, 5 iterations), with pre-reconstruction smoothing using a 3D Butterworth filter (cutoff: 0.45 cycles/cm; order 5), a 128 × 128 image matrix, a 1.23 zoom, and a pixel size of 3.9 mm. SPECT images were reconstructed using scatter correction (scatter energy window) and CT attenuation correction. CT and SPECT images were matched and fused into trans-axial images. A multimodality computer platform (Symbia net; Siemens) was used for image review and manipulation. Different 3D ROIs were drawn on the CT images with the software to delimit peripheral areas from proximal tracheobronchial areas of the lungs.

Results

Particle size distribution

As summarized in Table 1, the nebulizing systems led to different AMAD from sub-micron to micron sized particles: 2.8 μm (GSD of 3.2) for the Atomisor NL11, 550 nm (GSD of 2.1) for the Sidestream, and 230 nm (GSD of 1.6) for the Nanoneb. The geometric standard deviation clearly showed that the particle size distributions are broad and quite far from a monodisperse distribution. Thus the aerosol distribution can be characterized by the means of the [d16–d84] particle size range: [1 μm–9 μm] for Atomisor NL11, [0.25 μm–1 μm] for the modified Sidestream and [0.15 μm–0.5 μm] for the Nanoneb.

Table 1 Features of the inhaled aerosols to assess the impact of particle size on the regional distribution

Regional aerosol deposition by 2D gamma-imaging

The ROIs determined by 2D imaging after planar scintigraphies allowed to quantify the ET and TH deposited fractions (Table 2) but also an approximation of the exhaled fraction (Fig. 3) after the nasal administration of the polydisperse aerosol.

Table 2 Deposited fractions in extrathoracic (ET) and thoracic (TH) regions for the different types of polydisperse radioactive aerosols. Results are expressed in ratio (%) of the total deposited fraction (mean ± SD). Comparison of experimental results obtained with the chimeric model to baboon and human in vivo experimentations realized in previous studies [11, 17]
Fig. 3
figure 3

Respective percentages of exhaled fractions obtained for each [d16–d84] size range: [0.15 μm–0.5 μm], [0.25 μm–1 μm] and [1 μm–9 μm]. Results are expressed in ratio (%) of the nebulizer emitted fraction (mean ± SD, **p < 0,01)

Besides, the exhaled fraction decreased as the size of aerosol increased. More precisely, for the [0.15 μm–0.5 μm] aerosol particles, we noticed a very significant exhaled fraction (68 ± 12 %, expressed as a percentage of the emitted aerosol fraction by the nebulizer). In contrast, the exhaled fraction for the [0.25 μm–1 μm] and the [1 μm–9 μm] aerosol particles was quite similar (respectively 34 ± 11 % and 27 ± 9 %).

ET deposited fractions obtained for the three types of polydisperse aerosols were 11 ± 4 % for the [0.15 μm–0.5 μm] size range, 22 ± 5 % for [0.25 μm–1 μm] range and 65 ± 11 % for [1 μm–9 μm] range. In the same way, TH deposited fractions were 89 ± 4 % for the [0.15 μm–0.5 μm] size range, 78 ± 5 % for [0.25 μm–1 μm] aerosol and 35 ± 11 % for [1 μm–9 μm] aerosol. These data were compared to known in vivo data obtained in previous studies with baboons and humans (Table 2) using similar experimental conditions [11, 17].

Pulmonary deposition assessed by SPECT/CT

The 3D anatomical ROIs determined by CT scans with radioactivity counted by SPECT are shown in Fig. 4. With this methodology, the ROI outlines allowed the delimitation of the proximal and peripheral regions of the lungs. The proximal region corresponds to the trachea and the two principal bronchi. Representative SPECT-CT images of regional lung deposition obtained for the different polydisperse aerosols are shown in Fig. 4a. Interestingly, there were no significant differences among the three types of polydisperse radioactive aerosols [26]. Indeed, a large deposition of aerosols was clearly visible in the peripheral lung region whereas a low amount of radiolabeled airborne particles was deposited in the proximal tracheobronchial region (less than 5 % of the total aerosol deposition). The percentages of aerosol deposited in the lung (see ET fraction in Table 2) were normalized to 100 % to compare the proximal (central) versus peripheral regions of the lungs and are illustrated in Fig. 4b. The thoracic deposition percentage varies significantly in the model depending on the AMAD but whatever the thoracic deposition amount, this deposit is still peripheral.

Fig. 4
figure 4

a SPECT/CT imaging of the lung and delimitation of the central and peripheral regions of interest (red). b Percentage of deposition in the central and peripheral regions obtained for each [d16–d84] size range: [0.15 μm–0.5 μm], [0.25 μm–1 μm] and [1 μm–9 μm]

Discussion

In this study, the use of the chimeric model appeared useful in performing regional aerosol deposition cartographies for nasally administered aerosols. Moreover, the ET and TH deposited fractions for particles in the [1 μm–9 μm] and the [0.15 μm–0.5 μm] range were in acceptable agreement between inhalation experiments on baboons, healthy human volunteers and the chimeric model (i.e. respectively 72 ± 17 %, 72 ± 14 %, 65 ± 11 % for ET deposited fraction for the [1 μm–9 μm] size range) with non significant statistical differences (p = 0.4799). By contrast, significant differences (p < 0.0001) were noticed for the [0.25 μm–1 μm] aerosol (i.e. 22 ± 5 % in the chimeric model versus 49 ± 8 % in baboons for the ET deposited fraction).

Anatomically, the respiratory tracts of pigs are very similar to those of humans, with the same 23 branching divisions except for position of the tracheal bronchus, making pigs very good animal models of the human respiratory tract for decades [27]. Because its half-life after death is a few hours, surfactant is missing in the ex vivo porcine lung. This had to be taken into account in our experimental conditions. The simulated intra-pleural depressions necessary to inflate the lungs in the sealed enclosure were adjusted to compensate the absence of surfactant. As a result, the model requires twice the values of depression usually obtained for intrapleural depression during human breathing, which is consistent with the biomechanical role of the surfactant (approximately 50 % of the compliance) [28]. It has to be noted that the absence of blood circulation in this model impacts the compliance too.

Besides, limitations of inhalation experiments using the chimeric model have to be underlined which could explain differences observed for airborne particles in the [0.25 μm–1 μm] range.

On the one hand, the flap valve installed to simulate the vocal folds may filter larger particles much less or more effectively than real vocal folds depending on the occluded area and the flexibility. Nevertheless, the valve’s size and resistivity were carefully selected and this system has already been described and used in previous studies [13, 18]. Moreover previous studies using the proposed model were for gas ventilation (radioactive krypton) experiments without aerosols (i.e. particles or droplets suspended in the air). These points have to be taken into account for the extrathoracic/thoracic results reported in Table 2.

On the other hand, in the chimeric model, the inhalation mode and postural conditions vary from the compared studies (in vivo inhalation experiments with baboon and healthy human volunteer):

  • supine position with a nasal administration of the aerosol was used during the experiments with the chimeric model.

  • erect position with a nasal administration of the aerosol for the in vivo human experiments.

  • erect position with an oro/nasal administration for the in vivo baboon experiments.

For a given airborne particle size range and thus a specific deposition mechanism, this variation of inhalation mode and postural condition can play an important role and have a direct impact on comparisons between baboon, human and the chimeric model. Indeed, airborne particles can deposit by different ways in the respiratory tract, and it is possible to correlate the aerodynamic diameters to the deposition mechanisms in the respiratory airways. Inertial impaction, gravitational sedimentation, and turbulence-driven diffusion mainly govern the deposition of aerosol particles. Each deposition mechanism is directly related to aerodynamic particle size, airway geometry and particle velocity [2932]. For example, the respiratory deposition of particles in the [1 μm–9 μm], [0.15 μm–0.5 μm] and [0.25 μm–1 μm] size ranges are governed mainly by impaction, turbulence-driven diffusion and sedimentation mechanisms, respectively. Turbulence-driven diffusion mechanism is by random Brownian diffusion mainly in the pulmonary region. Some researchers have recently differentiated a turbulent diffusion mechanism as an addition to their impaction models in the upper airways [33].

It seems interesting to notice that the correlation of in vivo experiments performed in baboon and human compared to the ex vivo model is in good agreement for all airborne particle sizes, except those that have a mechanism of deposition by sedimentation mainly due to by gravity, i.e. the [0.25 μm–1 μm] range. Concerning the [1 μm–9 μm] aerosols, the deposition pattern observed in human, baboon and ex vivo model is very similar (around 70 % of deposition in the ENT region, p = 0,4799, no significant difference). These results confirm that there is no impact of the inhalation mode and the postural conditions when the deposition of aerosols is mainly governed by impaction. Similarly for the [0.15 μm–0.5 μm] aerosols, deposition is comparable between baboon and the ex vivo model (around 85 % of deposition in the thoracic region, p = 0,0623, no significant difference). These results confirm that there is a minimal impact of the inhalation mode and the postural conditions when the deposition of aerosol is mainly governed by diffusion. In contrast, disparate deposition results were observed for the [0.25 μm–1 μm] aerosols between the chimeric model and studies on baboon (p < 0,0001, significant difference). This result suggests that there is a high impact of the respiratory mode and the postural conditions when the deposition of aerosol is mainly governed by sedimentation.

To sum up, for sedimentation mechanism, the postural position (erect or supine) has a significant influence on the regional deposition whereas it is less important for the impaction and diffusion processes where the gravity does not play a major role [3437]. We assume that the regional aerosol deposition differences observed between the proposed ex vivo model and the in vivo experiments for the [0.25 μm–1 μm] aerosols can be mainly attributed to the influence of postural conditions. The limit of the ex vivo respiratory tract is the supine position during the inhalation experiment modifying the effects of gravity during the aerosolization compared to in vivo data obtained in the erect position. Thus, the chimeric model appears as quite reliable for predicting regional aerosol deposition for all aerosol size ranges, except for aerosol deposition governed by sedimentation as an inhalation experiment in erect position is needed. Technical solutions to perform the same experiments in erect position are under active consideration progress. The proposed ex vivo respiratory tract is an important tool to address the three R’s approach for aerosol deposition studies (Replacement: methods that do not employ animals; Reduction: methods that result in the use of fewer animals than existing methods; Refinement: methods or techniques that reduce pain, distress or discomfort to the animal), encouraging alternatives to in vivo animal testing.

Conclusion

This study presents original regional deposition data for submicronic and micronic particles in an ex vivo respiratory tract and compares these to in vivo data. Results clearly demonstrate a quite good agreement between the regional aerosol deposition obtained for [1 μm–9 μm] and [0.15 μm–0.5 μm] polydisperse aerosols compared to in vivo inhalation experiments in baboons, healthy human volunteers and the ex vivo model. However, a bias induced by the use of the ex vivo respiratory tract (with experiments currently performed in supine position, see Fig. 1) was noticed for the aerosols with a deposition mechanism governed by gravity (0.25 μm–1 μm range) in comparison with in vivo inhalation experiments performed in the erect position (0.25 μm–1 μm range). In conclusion, this study shows that the ex vivo respiratory tract proposed appears as quite reliable to assess aerosol deposition for a wide range of polydisperse aerosols (from sub-micron to micron sized particles), inhalation mode (nasal and oro/nasal administration) and postural positions (supine and erect positions). Thus, the methodology developed is useful to better understand the regional deposition of aerosols within the respiratory tract raising interesting ethical and cost issues, mainly to involve alternative methods as defined by the three R’s approach.

Abbreviations

3D OSEM, three-dimensional ordered subsets expectation-maximization; 99mTc-DTPA, Technetium 99 m (metastable) Diethylene Triamine Pentaacetic Acid; AMAD, activity median aerodynamic diameter; ELPI, electrical low-pressure impactor; ET, extrathoracic; GSD, geometric standard deviation; I/E ratio, inspiratory-to-expiratory time ratio; MBq, MegaBecquerel; ROI, region of interest; SD, standard deviation; SPECT, single-photon emission computed tomography; TH, thoracic; VT, tidal volume

References

  1. Kannan R, Guo P, Przekwas A. Particle transport in the human respiratory tract: formulation of a nodal inverse distance weighted Eulerian–Lagrangian transport and implementation of the Wind–Kessel algorithm for an oral delivery. Int J Numer Methods Biomed Eng. 2016;32.

  2. Löndahl J, Möller W, Pagels JH, Kreyling WG, Swietlicki E, Schmid O. Measurement techniques for respiratory tract deposition of airborne nanoparticles: a critical review. J Aerosol Med Pulm Drug Deliv. 2013;27:229–54.

    Article  PubMed  Google Scholar 

  3. Daigle CC, Chalupa DC, Gibb FR, Morrow PE, Oberdörster G, Utell MJ, et al. Ultrafine particle deposition in humans during rest and exercise. Inhal Toxicol. 2003;15:539–52.

    Article  CAS  PubMed  Google Scholar 

  4. Laube BL. In vivo measurements of aerosol dose and distribution: clinical relevance. J Aerosol Med Off J Int Soc Aerosols Med. 1996;9 Suppl 1:S77–91.

    Google Scholar 

  5. McKinney W, Jackson M, Sager TM, Reynolds JS, Chen BT, Afshari A, et al. Pulmonary and cardiovascular responses of rats to inhalation of a commercial antimicrobial spray containing titanium dioxide nanoparticles. Inhal Toxicol. 2012;24:447–57.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  6. Oosthuizen MA, Oberholzer HM, Scriba MR, van der Spuy WJ, Pretorius E. Evaluation of the morphological changes in the lungs of BALB/c mice after inhalation of spherical and rod-shaped titanium nanoparticles. Micron Oxf Engl. 2012;43:863–9.

    CAS  Google Scholar 

  7. Kuempel ED, Tran CL, Castranova V, Bailer AJ. Lung dosimetry and risk assessment of nanoparticles: evaluating and extending current models in rats and humans. Inhal Toxicol. 2006;18:717–24.

    Article  CAS  PubMed  Google Scholar 

  8. Awadalla M, Miyawaki S, Abou Alaiwa MH, Adam RJ, Bouzek DC, Michalski AS, et al. Early airway structural changes in cystic fibrosis pigs as a determinant of particle distribution and deposition. Ann Biomed Eng. 2014;42:915–27.

    Article  PubMed  Google Scholar 

  9. Ford WR, Blair AE, Evans RL, John E, Bugert JJ, Broadley KJ, et al. Human parainfluenza type 3 virus impairs the efficacy of glucocorticoids to limit allergy-induced pulmonary inflammation in guinea-pigs. Clin Sci Lond Engl 1979. 2013;125:471–82.

    CAS  Google Scholar 

  10. Manunta MDI, McAnulty RJ, McDowell A, Jin J, Ridout D, Fleming J, et al. Airway deposition of nebulized gene delivery nanocomplexes monitored by radioimaging agents. Am J Respir Cell Mol Biol. 2013;49:471–80.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Albuquerque-Silva I, Vecellio L, Durand M, Avet J, Le Pennec D, de Monte M, et al. Particle deposition in a child respiratory tract model: in vivo regional deposition of fine and ultrafine aerosols in baboons. PLoS One. 2014;9:e95456.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Durand M, Pourchez J, Louis B, Pouget JF, Isabey D, Coste A, et al. Plastinated nasal model: a new concept of anatomically realistic cast. Rhinology. 2011;49:30–6.

    PubMed  Google Scholar 

  13. Leclerc L, Pourchez J, Prevot N, Vecellio L, Le Guellec S, Cottier M, et al. Assessing sinus aerosol deposition: Benefits of SPECT–CT imaging. Int J Pharm. 2014;462:135–41.

    Article  CAS  PubMed  Google Scholar 

  14. Le Guellec S, Le Pennec D, Gatier S, Leclerc L, Cabrera M, Pourchez J, et al. Validation of anatomical models to study aerosol deposition in human nasal cavities. Pharm Res. 2014;31:228–37.

    Article  PubMed  Google Scholar 

  15. Durand M, Rusch P, Granjon D, Chantrel G, Prades JM, Dubois F, et al. Preliminary study of the deposition of aerosol in the maxillary sinuses using a plastinated model. J Aerosol Med Off J Int Soc Aerosols Med. 2001;14:83–93.

    Article  CAS  Google Scholar 

  16. Durand M, Pourchez J, Aubert G, Le Guellec S, Navarro L, Forest V, et al. Impact of acoustic airflow nebulization on intrasinus drug deposition of a human plastinated nasal cast: new insights into the mechanisms involved. Int J Pharm. 2011;421:63–71.

    Article  CAS  PubMed  Google Scholar 

  17. Durand M, Le Guellec S, Pourchez J, Dubois F, Aubert G, Chantrel G, et al. Sonic aerosol therapy to target maxillary sinuses. Eur Ann Otorhinolaryngol Head Neck Dis. 2012;129:244–50.

    Article  CAS  PubMed  Google Scholar 

  18. Leclerc L, Pourchez J, Aubert G, Leguellec S, Vecellio L, Cottier M, et al. Impact of Airborne Particle Size, Acoustic Airflow and Breathing Pattern on Delivery of Nebulized Antibiotic into the Maxillary Sinuses Using a Realistic Human Nasal Replica. Pharm Res. 2014;31:2335–43.

    Article  CAS  PubMed  Google Scholar 

  19. Leclerc L, Merhie AE, Navarro L, Prévôt N, Durand M, Pourchez J. Impact of acoustic airflow on intrasinus drug deposition: New insights into the vibrating mode and the optimal acoustic frequency to enhance the delivery of nebulized antibiotic. Int J Pharm. 2015;494:227–34.

    Article  CAS  PubMed  Google Scholar 

  20. Perinel S, Pourchez J, Leclerc L, Avet J, Durand M, Prévôt N, et al. An ex vivo chimeric human-porcine preclinical model for aerosol deposition studies. 2015. In Press

  21. Merhie AE, Navarro L, Delavenne X, Leclerc L, Pourchez J. A new Strategy to Improve Drug Delivery to the Maxillary Sinuses: The Frequency Sweep Acoustic Airflow. Pharm Res. 2016;33(5)1074–84.

    Article  PubMed  Google Scholar 

  22. Moller W, Saba GK, Haussinger K, Becker S, Keller M, Schuschnig U. Nasally inhaled pulsating aerosols: lung, sinus and nose deposition. Rhinology. 2011;49:286–91.

    CAS  PubMed  Google Scholar 

  23. Booker R. Interpretation and evaluation of pulmonary function tests. Nurs Stand R Coll Nurs G B 1987. 2009;23:46–56. quiz 58.

    Google Scholar 

  24. Criée CP, Sorichter S, Smith HJ, Kardos P, Merget R, Heise D, et al. Body plethysmography – Its principles and clinical use. Respir Med. 2011;105:959–71.

    Article  PubMed  Google Scholar 

  25. Flesch JD, Dine CJ. Lung volumes: Measurement, clinical use, and coding. CHEST J. 2012;142:506–10.

    Article  Google Scholar 

  26. Grgic B, Finlay WH, Burnell PKP, Heenan AF. In vitro intersubject and intrasubject deposition measurements in realistic mouth–throat geometries. J Aerosol Sci. 2004;35:1025–40.

    Article  CAS  Google Scholar 

  27. Ibrahim Z, Busch J, Awwad M, Wagner R, Wells K, Cooper DKC. Selected physiologic compatibilities and incompatibilities between human and porcine organ systems. Xenotransplantation. 2006;13:488–99.

    Article  PubMed  Google Scholar 

  28. West JB. L’essentiel sur la physiologie respiratoire. Essent. Sur Physiol. Respir. 6° édition. Paris, France: EDISEM-MALOINE; 2003.

  29. Clark AR. MDIs: Physics of Aerosol Formation. J Aerosol Med. 1996;9:S–19.

    Article  Google Scholar 

  30. Everard ML, Clark AR, Milner AD. Drug delivery from jet nebulisers. Arch Dis Child. 1992;67:586–91.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  31. Crowe CT, Schwarzkopf JD, Sommerfeld M, Tsuji Y. Multiphase Flows with Droplets and Particles, Second Edition. 2 edition. Paris, France: CRC Press; 2011.

  32. Crowe CT, Elger DF, Roberson JA. Engineering Fluid Mechanics. 8 edition. Paris, France: Wiley; 2006.

  33. Darquenne C, Fleming JS, Katz I, Martin AR, Schroeter J, Usmani OS, et al. Bridging the gap between science and clinical efficacy: physiology, imaging, and modeling of aerosols in the lung. J Aerosol Med Pulm Drug Deliv. 2016;29:107–26.

    Article  PubMed  Google Scholar 

  34. Sá RC, Zeman KL, Bennett WD, Prisk GK, Darquenne C. Effect of posture on regional deposition of coarse particles in the healthy human lung. J Aerosol Med Pulm Drug Deliv. 2015;28(6):423–31.

    Article  PubMed  Google Scholar 

  35. Clay MM, Clarke SW. Effect of nebulised aerosol size on lung deposition in patients with mild asthma. Thorax. 1987;42:190–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  36. Becquemin MH, Swift DL, Bouchikhi A, Roy M, Teillac A. Particle deposition and resistance in the noses of adults and children. Eur Respir J. 1991;4:694–702.

    CAS  PubMed  Google Scholar 

  37. Becquemin MH, Roy M, Robeau D, Bonnefous S, Piechowski J, Teillac A. Inhaled particle deposition and clearance from the normal respiratory tract. Respir Physiol. 1987;67:147–58.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to acknowledge the financial support of, Saint-Etienne Métropole and the Conseil Général de la Loire. The authors also acknowledge Gwendoline Sarry for technical assistance.

Authors’ contributions

JP, MD, SP, LL contributed to the conception and the design of research. SP, JP, NP, AD, LL materially participated in the research experiments analysis and interpretation of data. LL and JP were involved with drafting article. LL, JP, SP, NP materially participated in article preparation. JMV, MC and MD and all others were involved with reviewing the article. All authors have approved the final version of this article.

Competing interest

Marc Durand is a scientific consultant for DTF Medical society.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Lara Leclerc.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Perinel, S., Leclerc, L., Prévôt, N. et al. Micron-sized and submicron-sized aerosol deposition in a new ex vivo preclinical model. Respir Res 17, 78 (2016). https://doi.org/10.1186/s12931-016-0395-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12931-016-0395-7

Keywords