Normal spirometry equates to normal impulse oscillometry in healthy subjects

Background

Impulse oscillometry (IOS) is an effort independent test performed with tidal breathing which can ascertain the presence of small airways dysfunction by measuring the heterogeneity of resistance between 5 and 20 Hz (R5–R20) or the area under the reactance curve (AX). The article of Li et al. reported on impulse oscillometry (IOS) values in young non smoking symptomatic otherwise healthy subjects with preserved spirometry without a diagnosis of chronic respiratory disease [1]. In subjects who had forced expiratory flow rates < 65% predicted they reported poor area under curve values < 0.70 with associated low sensitivity and specificity values on the whole being less than 80%.

In our clinic we would consider pragmatic abnormal IOS values as R5 > 0.5 kPa/l.s, R5–R20 > 0.1 kPa/l/s and AX > 1.0 kPa/l, in terms of being indicative of a clinically relevant degree of airflow obstruction [2]. We would also like to point out that in persistent asthma patients with preserved spirometry who have FEV1 > 80% predicted, the presence of raised values for R5–R20 was associated with a greater likelihood of long term poor disease control [3]. Moreover small airways dysfunction as R5–R20 and AX is related to type 2 inflammation in asthma [4].

Li et al. have demonstrated that otherwise healthy individuals with symptoms who have abnormal forced expiratory flows also have normal IOS values. Moreover it is important to note that forced expiratory rates are rather volume dependent such that false positive low values are commonly seen when subjects do not exhale fully to residual volume. We would therefore have concerns that clinicians might end up erroneously interpreting what are essentially normal IOS values which would lead to unnecessary further investigation and cause potential unwarranted anxiety to patients with normal tests.

Conclusions

Impulse oscillometry should be interpreted in the light of abnormal values as being indicative of small airways dysfunction.

Availability of data and materials

Not applicable.

Abbreviations

R5:

Resistance at 5 Hz (total resistance)

R20:

Resistance at 20 Hz (central resistance)

R5–R20:

Resistance heterogeneity between 5 and 20 Hz (peripheral resistance)

AX:

Area under reactance curve

FEV1:

Forced expiratory volume in 1 s

IOS:

Impulse oscillometry

References

  1. Li LY, Yan TS, Yang J, Li YQ, Fu LX, Lan L, et al. Impulse oscillometry for detection of small airway dysfunction in subjects with chronic respiratory symptoms and preserved pulmonary function. Respir Res. 2021;22:68.

    Article  Google Scholar 

  2. Lipworth B, Manoharan A, Anderson W. Unlocking the quiet zone: the small airway asthma phenotype. Lancet Respir Med. 2014;2:497–506.

    Article  Google Scholar 

  3. Manoharan A, Anderson WJ, Lipworth J, Ibrahim I, Lipworth BJ. Small airway dysfunction is associated with poorer asthma control. EurRespir J. 2014;44:1353–5.

    Article  Google Scholar 

  4. Kuo CR, Jabbal S, Lipworth B. Is small airways dysfunction related to asthma control and type 2 inflammation? Ann Allergy Asthma Immunol. 2018;121:631–2.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

BJL and RC were responsible for writing the letter and read and approved the manuscript. All contributors who do not meet the criteria for authorship should be listed in an acknowledgements section. Both authors read and approved the final manuscript.

Corresponding author

Correspondence to Brian Lipworth.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

Drs. Lipworth and Chan have no conflicts of interest to report.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lipworth, B., Chan, R. Normal spirometry equates to normal impulse oscillometry in healthy subjects. Respir Res 22, 96 (2021). https://doi.org/10.1186/s12931-021-01693-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12931-021-01693-0