Skip to main content

Table 4 Estimated delivered and respirable doses upon nebulization of 200 mg of IgG. Breath simulation experiments and laser diffraction measurements were conducted upon nebulization of 2 ml (200 mg) of IgG to measure delivered and respirable doses (RD)

From: Topical application of nebulized human IgG, IgA and IgAM in the lungs of rats and non-human primates

Formulation

Nebulized dose (mg)

Residue (%)

Delivered dose (mg)

Delivered dose (%)

RD < 5 μm (mg)

RD < 5 μm (%)

IgG 10%

200

32.7 ± 8.10

94.06 ± 18.14

45.4 ± 9.00

72.24 ± 13.64

34.91 ± 6.78

IgG 5%

100

33 ± 9.10

49.36 ± 7.37

47.7 ± 7.10

34.36 ± 4.91

33.21 ± 4.74

IgA 5%

100

30.4 ± 6.90

48.41 ± 6.73

48.9± 6.90

35.33 ± 4.59

35.69 ± 4.63

IgAM 5%

100

30.8 ± 6.00

50.92 ± 6.32

51.4 ± 6.30

38.96 ± 5.37

39.34 ± 5.29