Figure 3

From: A comparison of biologically variable ventilation to recruitment manoeuvres in a porcine model of acute lung injury

Figure 3

Human Respiratory Rate Variability File. Human breathing variability file used for the study. This is the raw data from a spontaneously breathing healthy female volunteer. The mean rate was 13.4 ± 2.0 breaths/min (shown as the red line). There are 1587 breaths in this file. With BVV, the ventilator is configured as a volume divider at a fixed minute ventilation so that respiratory rate × tidal volume product is constant. Thus the breath-by-breath volume related to instantaneous respiratory rate obtained from sequentially reading the above file in any given experiment is obtained from the minute ventilation/[(instantaneous breath rate/13.4) × chosen mean rate]. Analysis reveals that these data have fractal characteristics.